Another source of guidance on stability testing is the World Health Organization, which has published
"Stability testing of active pharmaceutical ingredients and finished pharmaceutical products Annex 2" as part of their
Technical Report Series.
"The equipment used for the ongoing stability programme should be qualified and maintained following the general rules of EU GMP Volume 4, Chapter 3 (Premise and Equipment) and Annex 15 (Qualification and Validation)".
Stability Chambers & Redundant Monitoring SensorsLike other sensors, humidity sensors may drift over time. As drift occurs, control loops continue to respond to sensor inputs, so it's possible that everything may appear to be in control and within specification. Adding one or more redundant sensors reduces the risk of undetected sensor drift. It's best to detect out of specification conditions before stability studies are compromised. The solution is often to install redundant sensors to offset risks to product safety and efficacy.
Independent monitoring sensors need to be highly accurate, stable, and (ideally) equipped to send alerts outside certain thresholds. Most importantly, redundant sensors must be suitable for the intended conditions (e.g. 40°C / 75% humidity) and calibrated to ICH guidelines with a multi-point calibration.
Although sensor drift may occur too slowly to be detected, it can occur faster than most calibration cycles. Even small levels of contamination can cause significant drift. Calibrating once per year, or even every six months depending on how demanding your stability tests are, is expected.
How often do we need to ReQualify Stability Chambers?
Paul Daniel often receives questions about alarm limits and validation-mapping frequency for stability chambers.
Paul writes:For re-validation intervals, I think it's critical to consider that even when operated under a good change control program, things wear out. In the case of a stability chamber – motors wear, door seals warp, valves slowly clog, desiccant ages, and insulation hardens.
It falls to a process experts to determine how often to qualify their stability chambers. It's a question of risk tolerance, and it's predicated on the quality guidelines of individual companies, their history with auditors, and most importantly; the sensitivity of their processes.
Personally, I think auditors are interested in knowing revalidation intervals for stability chambers. But, I also think that if you have data that indicates the systems you use for stability testing are very stable, auditors would accept a longer frequency of remapping.



Upper air soundings provide the crucial data necessary for a complete atmospheric weather profile that leads to accurate forecasting. However…

Vaisala Automatic Weather Station AWS310 is a complete communication and weather observation solution that can be used as a standalone unit, or as…

Our next-generation, cyber secure and modular Maritime Automatic Weather Station AWS830 is engineered for harsh marine environments and compliant…

The Vaisala Automatic Weather Station AWS810 enables modern, high-quality surface weather observation networks anywhere in the world.

Vaisala Maritime Automatic Weather Station AWS830 delivers accurate, always‑on weather, environmental, oceanographic, and motion data for demanding…

Vaisala Beam Weather Station BWS500 is a compact, scalable station for weather, air quality and road condition data, from a single standalone station…

La Punto de Rocío y Temperatura DMP5 Vaisala DRYCAP® está diseñada para la medición de humedad en línea en aplicaciones de secado industrial con…

El Dr. Lauri Reuter es una experta en el futuro de los alimentos y la persona a quien preguntar si alguna vez te has preguntado qué vamos a comer en…

Las empresas industriales continúan tomando medidas para minimizar sus emisiones de carbono. Muchos están reduciendo el uso de combustibles fósiles y…

En el proceso de caustificación, los químicos pasivos se convierten en químicos de cocina activos para producir licor blanco. Se utiliza en el…
Un especialista responde con la solución, el instrumento y el servicio que corresponden.
Productos para cotizar
Todavía no hay productos. Se puede escribir igual: un especialista responde.