Marking of M2 Weights
According to OIML R111-1:2004, weights of class M2 from 1 g upwards are marked with their nominal value, but without unit information. To identify class M2, the class M2 is additionally applied.
For cylindrical weights with an adjustment cavity, the class marking is usually located on the knob or on the top of the weight. For block-shaped or other designs, the marking may be applied depending on the design and available space.
For differentiation, weights of the same nominal value that occur more than once in weight sets can be marked with dots or stars.
In addition, customer-specific user markings, such as characters, letters or numbers, can be applied. This allows the weights to be clearly identified as test or measuring equipment and better integrated into internal test equipment management.
These user markings can be specified individually when ordering.