Uma unidade da AKRIBIS One

PR1000-30-PSIA

PR1000 (PR1000-30-PSIA)

The PR1000 is a temperature, pressure and transient pressure data logger with an NPT fitting. Includes ISO/IEC 17025 Accredited Certificate.

Solicitar cotaçãoBaixar ficha técnica

PR1000 (PR1000-30-PSIA)

Key Features

  • Compact
  • Submersible
  • ¼ Inch NPT Fitting
  • Available in 10 Measurement Ranges
PR1000 (PR1000-30-PSIA)

Description

The PR1000 is a temperature, pressure and transient pressure data logger. This device is fully submersible and available in 10 models ranging from 30 to 5000 PSI.

Standing just under 4 inches tall and an inch in diameter, the PR1000 is compact for easy placement and features a ¼ inch NPT fitting for quick connection. The PR1000 can store more than 1.3 million readings with an internal temperature sensor that provides measurements from -40 °C to +80 °C (-40 °F to +176 °F) with an accuracy of ±2.0 °C.

The PR1000 is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

Note: The PR1000 replaces the PRTemp1000 and PRTrans1000.

PR1000 (PR1000-30-PSIA)

General

Memory2,794,837 readings
Reading Rate1 reading every second up to 1 reading every 5 minutes; Transient = 128Hz
Battery Life1 year; In Transient Mode: 10 Days (at 128Hz)
Material316 Stainless Steel/Radel
Dimensions3.6 in x 0.97 in dia. (91.3 mm x 24.6 mm dia.)
SubmersibleYes
Operating Environment-40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH
Required Interface PackageIFC400 or IFC406
ApprovalsCE

Temperature

Temperature SensorInternal semiconductor
Temperature Range-40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution1.0 °C
Temperature Calibrated Accuracy±2.0 °C

Pressure

Pressure Range0 to 30, 100, 300, 500 PSIA/PSIG; 1000 and 5000 PSIA
Pressure ResolutionRange: Resolution; 0 to 30 PSIA/PSIG: 0.0005 PSI; 100 PSIA/PSIG: 0.002 PSI; 300 PSIA/PSIG: 0.005 PSI; 500 PSIA/PSIG: 0.01 PSI; 1000 PSIA: 0.02 PSI; 5000 PSIA: 0.1 PSI
Pressure Calibrated AccuracyRange: Accuracy; 0 to 30 PSIA/PSIG: ±0.6 PSI; 100 PSIA/PSIG: ±2 PSI; 300 PSIA/PSIG: ±6 PSI; 500 PSIA/PSIG: ±10 PSI; 1000 PSIA: ±20 PSI; 5000 PSIA: ±100 PSI