Uma unidade da AKRIBIS One

PR1000Ex-300A

PR1000Ex (PR1000Ex-300A)

The PR1000Ex is a pressure and temperature data logger with NPT fitting that carries hazardous location certification (ATEX, IECEx). Includes ISO17025 accredited calibration certificate.

Solicitar cotaçãoBaixar ficha técnica

PR1000Ex (PR1000Ex-300A)

Key Features

  • ATEX Approved
  • Compact Design
  • ¼ Inch NPT Fitting
  • Available in 10 Measurement Ranges

Description

The PR1000Ex is a temperature, pressure, and transient pressure data logger that carries hazardous location certification (ATEX, IECEx).

  • Electrical Protection Concepts: IEC: 60079-11 Ex ia – Ex ic, Intrinsic Safety Zones 0-2
  • Equipment Protection Level: Ga – Gc, Zones 0-2
  • Gas Groups: IIC
  • Temperature Class: T4

Designed to fit in the palm of your hand, the PR1000Ex is only 3.6 inches tall and features a common ¼ in. NPT pressure port for easy installation. The PR1000Ex can store more than 1.3 million readings with an internal temperature sensor that provides measurements from -40 °C to +80 °C (-40 °F to +176 °F) with an accuracy of ±2.0 °C. To adapt to most applications, the PR1000Ex will be available in 10 different pressure ranges from 30 up to 5000 PSIA/PSIG.

The PR1000Ex is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

The PR1000Ex is certified in accordance with the latest issue of IECEx 60079-0, IECEx 60079-11, and Directive 2014/34/EU.

General

Memory2,794,837 readings
Reading Rate1 reading every second up to 1 reading every 5 minutes, Up to 128 Hz in transient mode
Battery Life1 year
Material316 Stainless Steel/Radel
Dimensions3.6 in x 0.97 in dia. (91.3 mm x 24.6 mm dia.)
SubmersibleYes
Operating Environment-40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH
Required Interface PackageIFC400/IFC406
ApprovalsCE, IS rated

Temperature

Temperature SensorInternal semiconductor
Temperature Range-40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution1.0 °C (1.8 °F)
Temperature Calibrated Accuracy±2.0 °C (3.6 °F) at 0 °C to +50 °C (+32 °F to +122 °F); ±4.0 °C (7.2 °F) at -40 °C to -1 °C (-40 °F to +30 °F); ±4.0 °C (7.2 °F) at +51 °C to +85 °C (+124 °F to +185 °F)

Pressure

Pressure SensorSemiconductor (strain gauge)
Pressure Range0-30, 0-100, 0-300, 0-500 PSIA/PSIG, 0-1000, 0-5000 PSIA
Pressure ResolutionSee Specifications Sheet
Pressure Calibrated AccuracyRange: Accuracy; 0 to 30 PSIA/PSIG: ±0.6 PSI; 100 PSIA/PSIG: ±2 PSI; 300 PSIA/PSIG: ±6 PSI; 500 PSIA/PSIG: ±10 PSI; 1000 PSIA: ±20 PSI; 5000 PSIA: ±100 PSI

Configurar o equipamento

Escolha modelo, sondas e acessórios. As combinações incompatíveis são bloqueadas na hora; as demais regras são verificadas ao adicionar o equipamento à cotação.

Documentos

  • IFC400

    IFC400

    Interface Cable Package for the HiTemp140 and PR140 Data Loggers.

  • IFC406

    IFC406

    IFC406 Multiplexer Data Logger Interface.

Escrever pelo WhatsApp