Gas Detectors E2611
2 × SPST relays, 2×4-20mA/0-10V, RS485 Modbus RTU, slotted ABS 85 × 85 × 37 mm, IP20
E2638
transmitter only (optional relays), 2 × 4-20mA/0-10V, RS485 Modbus RTU, grey ABS 90 × 145 × 50 mm, IP65
Solicitar cotizaciónDescargar ficha técnica
transmitter only (optional relays), 2 × 4-20mA/0-10V, RS485 Modbus RTU, grey ABS 90 × 145 × 50 mm, IP65

| Power supply | 12...36 VDC (default),; 24 VAC or 230 VAC as options |
|---|---|
| Digital interface | RS485, Modbus RTU protocol |
| Enclosure | grey ABS plastic, wall mount, IP65 |
| Model | Gas | Sensor type | Sampling method | Detection range | Resolution | Sensor lifetime | Analog outputs | Output relays (option) | Dimensions | Operating conditions | Default alarm set-points |
|---|---|---|---|---|---|---|---|---|---|---|---|
| E2638-C2H4 | ethylene C2H4 | electrochemical | diffusion | 0...10 ppm C2H4 0...200 ppm C2H4 0...1500 ppm C2H4 | 0,1 / 1 / 1 ppm C2H4 | 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), | H90 × W145 × D55 mm | -40...+50 °C, 15...90 %RH with humidity compensation up to 100 %RH | |
| E2638-CH4 | methane CH4 | metal-oxide semiconductor (MOS) pellistor (catalytic bead) on demand | diffusion | 0...100% LEL CH4 | 0,1 / 1 %LEL CH4 | > 5 years (MOS) > 3 years (pellistor) | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -40...+70 °C, 15...95 %RH (MOS) -10...+50 °C, 0...95 %RH (pellistor) | 20% LEL / 50% LEL CH4 |
| E2638-CL2 | chlorine Cl2 | electrochemical | diffusion | 0...10 ppm Cl2 | 0,1 ppm Cl2 | 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -20...+50 °C, 15...90 %RH | 0,5 ppm / 1 ppm Cl2 |
| E2638-CO | Carbon Monoxide CO | electrochemical | Diffusion | 0...300 ppm CO 0...1000 ppm CO | 1 ppm CO | > 10 years | 4-20 mA / 0-10 V, user settable | H90 × W145 × D55 mm | -20...+50 °C, 15...90 %RH | ||
| E2638-CO2 | carbon dioxide CO2 | photoacoustic optical NDIR | diffusion | 0...10 000 ppm (10K) CO2 0...40 000 ppm (40K) CO2 0...5 %vol (50KN) CO2 up to 100% CO2 on request | 1 ppm for 10K 10 ppm for 40K 0.01%vol for 50KN | up to 10 years for 10K/40K up to 5 years for 50KN | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), | H90 × W145 × D55 mm | -40...+60 °C, 0...100 %RH | |
| E2638-ETO | ethylene oxide C2H4O | electrochemical | diffusion | 0...20 ppm C2H4O 0...100ppm C2H4O | 0,1 ppm C2H4O for 20 ppm range 1 ppm C2H4O for 100 ppm range | 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -20...+50 °C, 15...90 %RH | |
| E2638-H2S | hydrogen sulfide H2S | electrochemical | diffusion | 0...100 ppm H2S 0...2000 ppm H2S other ranges on request | 0.1 / 1 ppm H2S | > 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -30...+50 °C, 15...90 %RH | 10 ppm / 20 ppm H2S |
| E2638-HFC | R-32, R-134a, R-404A, R-407C, R-410A, R-449A, R-454A, R-454B, R-454C, R-507, R-513A, R-1234ze, etc. | metal-oxide semiconductor optical NDIR on demand | diffusion | 0...1000 ppm halocarbon gas 0...2500 ppm halocarbon gas | 1 ppm halocarbon gas | > 5 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -30...+60 °C, < 95 %RH | minimum 250 ppm |
| E2638-LEL | methane CH4 | metal-oxide semiconductor (MOS) pellistor (catalytic bead) on demand | diffusion | 0...100% LEL | 0,1 / 1 % LEL | > 5 years (MOS) > 3 years (pellistor) | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -40...+70 °C, 15...95 %RH (MOS) -10...+50 °C, 0...95 %RH (pellistor) | 20% LEL / 50% LEL CH4 |
| E2638-N2O | nitrous oxide N2O | optical NDIR | diffusion | 0...1000 ppm N2O | 1 ppm N2O | up to 5 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), | H90 × W145 × D55 mm | -20...+50 °C, 0...95 %RH | |
| E2638-NH3-E | ammonia NH3 | electrochemical | diffusion | 0...100 ppm NH3 0...300 ppm NH3 0...1000 ppm NH3 0...5 000 ppm NH3 up to 10000 ppm on request | 1 ppm NH3 | > 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -40...+50 °C, -30...+50°C (for 0...5 000 ppm), 15...90 %RH | 25 ppm / 35 ppm NH3 |
| E2638-NH3-P | Ammonia NH3 | Pellistor (catalytic bead) | Diffusion | 0...15% vol. (0…100% LEL) | 0.1% vol. or 1% LEL | > 3 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -10...+50 °C, 0...95% RH | |
| E2638-NO | nitric oxide NO | electrochemical | diffusion | 0...250 ppm NO | 1 ppm NO | > 12 months | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -30...+50 °C, 15...90 %RH | 10 ppm / 25 ppm NO |
| E2638-NO2 | nitrogen dioxide NO2 | electrochemical | diffusion | 0...20 ppm NO2 0...200 ppm NO2 | 0,1 ppm NO2 for 20 ppm range 1 ppm NO2 for 200 ppm range | > 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -20...+50 °C, 15...90 %RH | 3 ppm / 6 ppm NO2 |
| E2638-O2 | oxygen O2 | electrochemical | diffusion | 0...25 %vol O2 | 0,01 %vol O2 | > 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), | H90 × W145 × D55 mm | -30...+50 °C, 15...90 %RH | 19,5 / 18,0 %vol O2 |
| E2638-O3 | ozone O3 | electrochemical | diffusion | 0...5 ppm, 0...1 ppm | 0,01 ppm | > 2 years (0...5 ppm) >18 months (0...1 ppm) | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -40...+50 °C (0...5 ppm), -20...+40 °C (0...1 ppm), 15...90 %RH | |
| E2638-PID | isobutylene | Photo Ionization Detector | diffusion | 0...40 ppm isobutylene 0...200 ppm isobutylene | 0,01 ppm C4H8 for 40 ppm range 1 ppm C4H8 for 200 ppm range | 5 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -40...+65 °C, < 95 %RH | |
| E2638-SF6 | sulfur hexafluoride SF6 | optical NDIR | diffusion | 0...1000 ppm SF6 0...2000 ppm SF6 | 1 ppm | ≥ 5 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), | H90 × W145 × D55 mm | -20...+50 °C, 15...95 %RH | |
| E2638-SO2 | sulfur dioxide SO2 | electrochemical | diffusion | 0...50 ppm SO2 0...2000 ppm SO2 | 0,1 ppm SO2 for 50 ppm range 1 ppm SO2 for 2000 ppm range | > 2 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90× W145 × D55 mm | -30...+50 °C, 15...90 %RH | |
| E2638-VOC | toluene | metal-oxide semiconductor | diffusion | 0...100% LEL (flammable range) 0...1000 ppm (toxic range) | 0,1% LEL for LEL range 1 ppm for ppm range | > 5 years | 2 × 4-20 mA / 0-10 V, user settable | 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max | H90 × W145 × D55 mm | -40…+70 °C, <95 %RH (non-condensing) | 20% LEL / 50% LEL 100 ppm / 300 ppm |
Elija modelo, sondas y accesorios. Las combinaciones incompatibles se bloquean solas; el resto de las reglas se verifica al agregar el equipo a la cotización.
Cargando las opciones…
Se puede pegar un código y la configuración se arma sola.
Se piden por separado: no hace falta configurar el equipo.

2 × SPST relays, 2×4-20mA/0-10V, RS485 Modbus RTU, slotted ABS 85 × 85 × 37 mm, IP20

2 × SPDT relays, slotted ABS 85 × 85 × 37 mm, IP20, LEDs, buzzer, 24 VDC / 24 VAC / 230 VAC


transmitter only, 2×4-20mA/0-10V, RS485 Modbus RTU, grey ABS 80 × 82 × 55 mm, IP65

2 x SPDT relays, ABS plastic 90 x 145 × 55 mm, IP65, LEDs, buzzer, 24 VDC / 24 VAC / 230 VAC

transmitter only, 2 x 4-20 mA / 0-10 V, RS485 Modbus RTU, grey ABS 90 × 145 × 55 mm, IP65

transmitter only (optional relays), 2 × 4-20mA/0-10V, RS485 Modbus RTU, grey die-cast aluminium 120 × 125 × 57 mm, IP66

E2639 is a member of the new-generation PluraSens® family of multifunctional measurement instruments.

The instrument utilizes novel fully calibrated and temperature compensated gas sensors with excellent repeatability, stability, and long lifetime.

Instrumentos para cotizar
Todavía no hay instrumentos. Se puede escribir igual: un especialista responde.