Medusa-R-999
Medusa R Model 999
The Medusa R Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range 30°C to 550°C.
Solicitar cotizaciónDescargar ficha técnica
Medusa-R-999
The Medusa R Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range 30°C to 550°C.
Solicitar cotizaciónDescargar ficha técnica
It is suitable for use as a radiation source for infrared thermometers. The cavity is 45 x 285mm deep and suitable for use with the larger Isotech fixed point cells. Laboratory performance and low uncertainty calibrations are ensured by the combination by high emissivity and excellent temperature uniformity.
The digital temperature controller allows the block temperature to bet set to any value from 30°C to 550°C. Uniformity of the large block is ensured by using distributed heating technology.
For the smallest of uncertainties the Medusa R may be used with Isotech ITS-90 Fixed Point Cells, Indium 156.5985°C, Tin 231.928°C and Zinc 419.527°C. The cells are provided with a certificate of metal purity.

| Model | 999 |
|---|---|
| Temperature Range | 30°C to 550.0°C |
| Emissivity | Greater than 0.995 |
| Stability | ±0.1°C |
| Display Resolution | 0.01°C to 99.99; 0.1°C from 100 to 550 |
| Heating Time | 45 minutes |
| Aperture Diameter | 45mm |
| Cavity Depth | 285mm |
| PC Interface | Included |
| Power | 1000 Watts typical |
| Voltage | 100-130 or 208-240 Vac 50/60Hz |
| Dimensions | Height 480mm Width 425mm Depth 260mm |
| Weight | 17kg |
| Options | Fixed Point Cells: Indium Large Primary Cell – 998-06-00A. Tin Hockey Large Primary Cell – 998-06-00B. Zinc Large Primary Cell 998-06-00C |
| How to order | Model 999 Medusa R – State supply voltage. Please state any special calibration requirement. |


When the high accuracy of the Gemini R is not necessary this product offers a cost effective solution for the calibration and testing of infrared…

The higher temperature Isotech Blackbody Fixed Point Cells consist of high purity metals within a graphite body. Graphite reacts with air to form…

Isotech first introduced the Model 988 to assist in relation to the SARS outbreak during the early 2000s. Used to increase the accuracy of thermal…

Addressing Calibration Challenges Traditional calibration methods using stirred water baths are cumbersome, expensive, and not easily portable…

The Gemini R 550 Portable Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range 30°C to…

Primary Standard cells fit into the Medusa and Oberon, covering from Indium to Copper Points. These cylindrical cells are 42mm diameter and have a…
Instrumentos para cotizar