Mercury Triple Point Apparatus
The Isotech Model ITL-M-17725 apparatus is a self-contained, mechanically-refrigerated, system with a main well to house one mercury cell and two…
POTTS-Medusa-Series
Need advice on which model? Dry Block Selection Guide
Solicitar cotaçãoBaixar ficha técnica

Isotech have a wide range of Dry Blocks to suit probes requiring a large immersion depth. These products feature large and deep calibration volumes. As such they are less portable than the earlier Dry Blocks, but have higher capacities and retain outstanding temperature uniformity, this uniformity is so good that these larger products are also apparatus for Secondary Laboratories to realize the Fixed Points of ITS-90.
Medusa 510 has a maximum operating temperature of 550°C. In addition to the main heating zone there are additional top and bottom heaters which compensate for the end losses creating a constant temperature zone across the well.
For Comparison Calibration the Medusa should be used with an insert, the standard insert has six 8mm pockets 250mm deep. Also available is an insert 44mm diameter x 170mm deep which is suspended from the top of the block so that the height is user adjustable. For flexibility the Medusa can also be used with accessories for infrared thermometers and surface sensors. The Medusa is available in two models, the BASIC (B) and the SITE (S). The B model includes a sophisticated temperature controller with a dual display for Set Temperature and Dry Block Temperature.

The S model includes a built-in digital thermometer to which an external standard thermometer can be connected giving greater accuracy, eliminating temperature gradient and loading errors. Also included in the site model is a timer which can set the bath between two temperatures, and automate ITS-90 fixed point operation. For Surface Sensor and Blackbody use an external thermometer is recommended. For laboratory accuracy the Medusa can be used with a high-end temperature indicator such as an Isotech TTI model.
Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature. Increased resolution of ±0.01 available throughout the range via the PC interface and from 0.01 to +99.99 locally on the autoranging front display. The controller features multi-point block to display correction giving good absolute accuracy.
The S model has universal sensor input allowing Platinum Resistance Thermometers, Thermocouples (types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including 4-20mA current transmitters to be displayed on the in-built indicator. The indicator can be programmed with up to five calibration points to provide high accuracy digital probe matching. The indicator and controller are both addressable over the communications link.
| Model | 510 Medusa 511 Medusa 3 |
|---|---|
| Temperature Range | 30°C to 550°C (510) and 50°C to 700°C (511) |
| Absolute Stability: | over 30 minutes: |
| Metal Block Bath | ±0.03°C |
| BlackBody Source | ±0.1°C |
| Surface Sensor Calibrator | ±0.5°C |
| ITS-90 Fixed Point | ±0.001°C |
| Computer Interface | Included with software |
| Cools from | 550°C to 30°C in 5 hours |
| Heats from | 30°C to 550°C in 90 minutes |
| Uncertainties | Refer to uncertainties graph |
| Calibration Volume | 45mm diameter by 285mm deep |
| Standard Insert | Six 8mm pockets all 250mm deep |
| Display Resolution | 0.01 to 99.99, 0.1 to 100.0 to 650.0. PC can display 0.1 across whole range with the software included. |
| Indicator Units | °C, °F, K |
| Power | 108-130V or 208-240V 50/60Hz. 510: 1000 Watts 511: 1800 Watts |
| Overall Dimensions | Height 430mm, Width 310mm, Depth 300mm, Weight 510: 17kg 511:25kg |
| How to order | 510 Medusa or 511 Medusa 3. Please specify model type, voltage and options required. |
Escolha modelo, sondas e acessórios. As combinações incompatíveis são bloqueadas na hora; as demais regras são verificadas ao adicionar o equipamento à cotação.
Carregando as opções…
É possível colar um código e a configuração se monta sozinha.
São cotados à parte: não é preciso configurar o equipamento.

The Isotech Model ITL-M-17725 apparatus is a self-contained, mechanically-refrigerated, system with a main well to house one mercury cell and two…

This dedicated apparatus has more than 35 years of proven use and is widely used by the worlds’ leading NMIs. In a recent International study cells…

The Isotech Argon Triple Point Apparatus has been upgraded for greater automation, efficiency, and ease of use in realising the argon triple point.

The multi-functions of the Europa-6 is a unique award winning concept from Isotech.

Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…

The Annealing Furnace, model 414, is designed to heat, anneal and cool Standard Platinum Resistance Thermometers (SPRTs) prior to calibration. The…

Isotech metrology furnaces have more than 35 years of proven use and are widely used by the worlds’ leading NMIs. For the optimal use of fixed point…

Isotech has produced over 200 high temperature thermometers which have been sold world-wide for use up to the silver point. As a consequence of our…
Instrumentos para cotar
Ainda não há instrumentos. É possível escrever mesmo assim: um especialista responde.