Model 471 Simple Argon Apparatus
The Isotech Argon Triple Point Apparatus has been upgraded for greater automation, efficiency, and ease of use in realising the argon triple point.
Slim-Fixed-Point-Cells
The Isotech B8 30 130 Cell is small enough to fit into portable Dry Blocks such as the Venus and Europa models. For the larger blocks like the Oceanus Isotech recommend the B12 46 210 with its increased immersion depth.
Solicitar cotaçãoBaixar ficha técnica

Water
Gallium and Mercury Cells
Like the B8 Cell the Slim Gallium Cell can be used in portable blocks like the Venus and Europa, or in stirred liquid baths. The Hydra and Orion have accessories available to support the cells.
Indium to Copper
Isotech’s Slim Cells have been in constant use since their introduction in 1990. The cells have always been made from the highest quality graphite and 6N (99.9999%) pure metals. After further investment in the lab, and gaining smaller uncertainties from UKAS, we reviewed and further refined our range of metal clad cells to give better accuracy and performance. The new professional ranges of cells have more metal inside providing an active immersion depth in the metal of 160mm.

Metal Clad
Isotech produced the first metal clad cells in 1990 and have much experience in the manufacture and calibration of high quality proven metal clad cells. Metal clad cells are recommended for all points from Indium to Aluminium.
Quartz Clad
These cells are recommended for Silver and Copper points, whilst available for the lower temperature points the metal clad versions are recommended as they are more robust, have the same performance and are more cost effective.
Equipment for Slim Cells
The Slim Cells can be used in the same apparatus as the larger cells, and the greater immersion depth will give the lowest uncertainties.

| Cells – Temperature | Uncertainty | Additional Uncertainty | Model Metal Clad | Model |
|---|---|---|---|---|
| Water – 0.01°C | ±0.5mk | ±0.3mk | N/A | B8 30 130 |
| Water – 0.01°C | ±0.1mk | ±0.3mk | N/A | B12 40 210 |
| Water – 0.01°C | ±0.1mk | ±0.3mk | N/A | B12 46 210 |
| Gallium – 29.7646°C | ±0.5mk | ±0.3mk | 17401M | N/A |
| Mercury – -38.8344°C | ±0.5mk | ±0.1mk | 17724M | N/A |
| Higher Temperature | Quartz Clad | |||
| Indium – 156.5985°C | ±0.6mk | ±0.6mk | 17668ML | 17668QS |
| Tin – 231.928°C | ±0.8mk | ±0.8mk | 17669ML | 17669QS |
| Zinc – 419.527°C | ±1mk | ±1.5mk | 17671ML | 17671QS |
| Aluminium – 660.323°C | ±2mk | ±3mk | 17672ML | 17672QS |
| Silver – 961.78°C | ±10mk | N/A | 17673QS | |
| Copper – 1084.62°C | N/A | 17674QS |

The Isotech Argon Triple Point Apparatus has been upgraded for greater automation, efficiency, and ease of use in realising the argon triple point.

The multi-functions of the Europa-6 is a unique award winning concept from Isotech.

Model 426 is for Aluminium, Silver, Gold or Copper slim fixed point cells as well as for comparison calibration. Heat pipes provide the ideal…

Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…

Isotech metrology furnaces have more than 35 years of proven use and are widely used by the worlds’ leading NMIs. For the optimal use of fixed point…

Instrumentos para cotar