Uma unidade da AKRIBIS One

POTTS-Oceanus-6-580

POTTS – Oceanus-6 Model 580

Need advice on which model? Dry Block Selection Guide

Solicitar cotaçãoBaixar ficha técnica

POTTS – Oceanus-6 Model 580

Features

  • Use with Water and Gallium Fixed Points
  • Can also be used as a Dry Block, Liquid Bath
POTTS – Oceanus-6 Model 580

Description

The multi-functions of the Oceanus-6 is a unique concept from Isotech, so unique that it is currently the subject of a patent application.

The Oceanus-6 is a complete calibration laboratory working over the temperature range 45°C below ambient temperature to +140°C. It permits the calibration of temperature sensors absolutely (at fixed points of the ITS-90 scale) or by comparison to a reference standard.

It will calibrate both contact and non-contact thermometers such as optical pyrometers, surface sensors, liquid in glass thermometers as well as thermistors, thermocouples and resistance thermometers, whether they are short, long or odd shaped. Further it permits maintenance of reference standards by confirming the Ice or Water triple point at regular intervals.

The Oceanus-6 has a calibration volume of 52mm diameter and 300mm deep and is supplied with the very latest technology digital indicator, timer and controller making the Oceanus-6 a complete self-contained calibration laboratory.

The Oceanus-6 offers unprecedented accuracies of ±0.0002°C (2 Sigma) at the Water triple point and the Gallium melt temperature of 29.7646°C and up to ±0.005°C in the stirred liquid bath option (by comparison).

POTTS – Oceanus-6 Model 580

Specifications

Model580
Temperature Range45°C below ambient to +140°C
Absolute Minimum Temperature-45°C
Absolute Stabilities:over 30 minutes:
Metal Block Bath±0.03°C
Stirred Liquid Bath±0.025°C
Ice/Water Bath±0.001°C
BlackBody Source±0.3°C
Surface Sensor Calibrator±0.5°C
ITS-90 Fixed Point±0.0002°C
Heating/CoolingSee graph
Stabilisation Time10 minutes
Calibration Volume52mm x 300mm deep
Uniformity±0.018°C
Controller Resolution0.1 to 0.01 (4 digit display)
Indicator Resolution0.1 to 0.01 (4 digit display)
Indicator Units°C, °F, K
CommunicationsSupplied as standard with serial interface, PC adaptor cable and Cal Notepad.
Power300 Watts, 108-130 or 208-240Vac, 50/60Hz
Dimensions430mm Height, 310mm Width, Depth 300mm, Weight 17kg
How to order580 Oceanus-6 POTTS. Please specify voltage required.

Documentos

  • Model-465

    Three Zone High Temperature Furnace 465

    Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…

  • Model-17402

    Gallium Fixed Point Apparatus

    Model 17402B Gallium Apparatus is not an adaptation from general purpose equipment, but is designed specifically to realize and maintain the Gallium…

  • Gold-Platinum-Thermocouple

    Gold / Platinum Thermocouple

    Since 1995 Isotech have been producing various designs of special Au/Pt, Pt/Pd, Pd/Au thermocouples for researchers. From our experience we can now…

  • ITS-90-Optimal-Fixed-Point-Cells

    ITS-90 Optimal Fixed Point Cells

    Isotech Ultra Pure-Metal Freezing Point Cells are specifically designed to realize the liquid-solid equilibrium temperatures of high-purity metal…

  • SPRT-670

    Isotech Model 670 SPRT

    The 670SQ is our latest thermometer, to be specifically designed to give optimum performance up to the aluminium point. Its construction permits the…

  • POTTS-Europa-6-520

    POTTS – Europa-6 Model 520

    The multi-functions of the Europa-6 is a unique award winning concept from Isotech.

  • Slim-Fixed-Point-Cells

    Slim Fixed Point Cells

    The Isotech B8 30 130 Cell is small enough to fit into portable Dry Blocks such as the Venus and Europa models. For the larger blocks like the…