Medidor de punto de rocío DMT132
para secadores refrigerantes
Nota técnica
When people think about dew point measurement, they often picture a straightforward industrial application. But in reality, some of the most interesting questions come from environments where even tiny amounts of moisture can create major problems.
We hear from researchers building custom test setups, engineers running ultra-dry inert gas systems, compressed air specialists verifying air quality, and manufacturers protecting sensitive materials from oxidation or contamination. And while the applications are very different, the questions are often surprisingly similar:
This FAQ brings together real questions we receive about the Vaisala DMT153 from demanding applications across research, aerospace, pharmaceuticals, glove boxes, compressed air, and high-purity gas systems.
If you are working with critical dry conditions, hopefully these answers make your next project a little easier.
A: You can read dew/frost point, dew point, water concentration (ppmv/ppbv), and water mass fraction (ppmw) via analog outputs or Modbus RTU. This makes integration into lab DAQ systems straightforward.
A: Standard delivery includes a traceable ISO9001 calibration certificate. You can follow normal Vaisala service routes for periodic factory calibration or use field checks via Insight/Indigo80 handheld for interim verification. We also offer accredited ISO17025 calibration certificates with custom points.
A: DMT153 is a robust industrial transmitter, and it complies with IEC/EN 61326 1 for industrial EMC and CISPR 11 / EN 55011 Class B. It has no explicit MIL STD qualification.
A: The DMT153 operating temperature is −40 … +70 °C and pressure 0–50 bar absolute. The sensor is designed to withstand fluctuating temperature and pressure in real processes, not only “ideal lab” conditions, which is important in environmental test chambers and similar setups.
A: Vaisala can customize and label DMT153 for OEM use depending on the customer’s needs. This requires continuous, reasonable sales volumes. Please ask for details.
A: Operating voltage is 11–28 V DC in digital mode and 15–28 V DC when using analog outputs; typical current is about 20 mA plus load. During self-diagnostics max. 220 mA (pulsed) that the power supply must always offer.
A: It supports monitoring of Class 1 compressed air as per ISO 8573‑1:2010
A: Suitable locations include ducts and pipes; various sampling cells with Swagelok/VCR/quick connectors and coil options are available for bypass or 2‑pressure installations. A bypass with a small continuous flow through a sampling cell is usually preferred for high‑pressure compressed air, allowing service without venting the main line and protecting the sensor from particles.
A: Measured dew/frost point temperature does not require pressure compensation. When using ppm/ppb outputs (water concentration) or water mass-fraction, please make sure the pressure setting for DMT153 is representative of the actual measurement environment.
A: Materials in contact with gas are stainless steel (AISI 316L, AISI 303 filter) with a stainless mesh filter, suitable for typical high‑purity lines.
A: You can mount the probe directly through a wall with threaded fitting or use a small sampling loop if preferred (e.g. to keep electronics outside a controlled atmosphere). Typical installation is into air/gas inlet/outlets.
A: When connected to an Indigo transmitter you can use local relays or additional analog channels for alarm thresholds, while in standalone mode you can use the built‑in analog channels and Modbus data to trigger alarms in your own control system.
A: Each unit comes with a traceable dew point calibration certificate, which supports validation and audit needs.
A: DMT153 complies with relevant EMC and RoHS directives. Vaisala can also provide material declaration of the product wetted parts.
A: Occasional exposure of the sensor to ambient humidity (for example, during installation or service) is acceptable and the sensor tolerates condensation. The purge function is specifically designed to recover performance after such events.
A: Choose a point in the line or chamber where the gas is well mixed and flowing. When using a sampling cell, maintain a continuous flow (>1 l/min) so the cell is constantly refreshed and does not act as a stagnant volume. Eliminate leaks between the process and sensor.

para secadores refrigerantes

para aplicaciones de fabricantes de equipos originales

Vaisala Weather Transmitter WXT530 series is a unique collection of cost-effective sensors that accurately and reliably measure weather conditions…

Transmisores de ±1,5 % HR para HVAC exigentes, con certificado

Para usar en las zonas con clasificación hasta 0 y 20

Transmisores de pared de ±1,7% HR con certificado para HVAC exigentes.

Qué son la humedad relativa, el punto de rocío, la humedad absoluta y la razón de mezcla, cómo la temperatura y la presión distorsionan la lectura, y…

Cómo se comporta la humedad en condiciones reales, los tres problemas de medición más frecuentes, y dos escenarios prácticos —aire calentado y aire…
Cuándo medir humedad relativa y cuándo punto de rocío, qué exige la alta y la baja humedad, cómo influyen temperatura y presión, y cuándo conviene…

La climatización representa hasta el 40 % de la demanda energética de un centro de datos. Esta guía de Vaisala explica cómo la exactitud de medición…
Un especialista responde con la solución, el instrumento y el servicio que corresponden.
Instrumentos para cotizar
Todavía no hay instrumentos. Se puede escribir igual: un especialista responde.