Current X16-3A
CurrentX (Current X16-3A)
The CurrentX Series includes 4, 8, 12 and 16-channel low-level DC current data loggers. Includes standard calibration certificate.
Solicitar cotizaciónDescargar ficha técnica
Current X16-3A
The CurrentX Series includes 4, 8, 12 and 16-channel low-level DC current data loggers. Includes standard calibration certificate.
Solicitar cotizaciónDescargar ficha técnica

The CurrentX Series includes the CurrentX4, CurrentX8, CurrentX12 and CurrentX16 low-level DC current data loggers and are available in three ranges: 30 mA, 160 mA, and 3 A.
In addition to having a large memory capacity, the CurrentX4 gives users the option to enable or disable channels to further maximize memory capacity. There’s no interface necessary, the CurrentX Series uses standard USB-A to Micro USB cable.
Using the MadgeTech Software, starting, stopping and downloading from the CurrentX Series data loggers are simple and easy. Graphical, tabular and statistical data is provided for analysis and data can be viewed in multiple units, using the Engineering Units function. The data can also be automatically exported to Excel® for further calculations.

| Memory | 4-channel: 1,048,064 readings; 8-channel: 524,032 readings; 12-channel: 349,354 readings; 16-channel: 262,016 readings |
|---|---|
| Reading Rate | 4 Hz to 1 reading every 24 hours |
| Channels | 4, 8, 12, 16 |
| LED Indicators | 1 per channel and 2 status LEDs |
| Battery Life | 18 months |
| Material | Black anodized aluminum |
| Dimensions | 4-channel: 2.70 in x 7.25 in x 1.22 in (65.6 mm x 184.2 mm x 31.0 mm); 8-channel: 2.70 in x 7.25 in x 1.22 in (65.6 mm x 184.2 mm x 31.0 mm); 12-channel: 2.70 in x 7.25 in x 1.68 in (65.6 mm x 184.2 mm x 42.7 mm); 16-channel: 2.70 in x 7.25 in x 2.14 in (65.6 mm x 184.2 mm x 54.4 mm) |
| Operating Environment | -20 °C to +60 °C (-4 °F to +140 °F), 0 %RH to 95 %RH non-condensing |
| Current Range | -2 mA to +30 mA, ±160 mA, ±3 A |
|---|---|
| Max Voltage: Inputs & Ground | 3 V; (Common mode voltage must be less than 3 volts. All inputs must be within 3 volts of all other inputs.) |
| Current Resolution | 30mA: 0.0005 mA; 160mA: 0.005 mA; 3A: 0.001 A |
| Current Calibrate Accuracy | 30mA: ±0.016 mA; 160mA: ±0.16 mA; 3A: ±0.009 A |
| Current Input Impedance | 30mA: 10 Ω; 160mA: 1 Ω; 3A: 0.1 Ω |