MagMount-1
Magnetic Holder for use with the PR140, PRTemp140, HiTemp140, Temp1000, and RHTemp1000 Series Data Loggers.
Temp1000IS-2
The Temp1000IS is an intrinsically safe, temperature data logger. Includes ISO17025 accredited calibration certificate.
Solicitar cotizaciónDescargar ficha técnica

The Temp1000IS is a compact, temperature data logger that is SGS approved as intrinsically safe for hazardous environments. The Temp1000IS is available with either a flush bottom or key ring end cap.
This device measures and records temperatures from -40 °C to +80 °C (-40 °F to +176 °F) with an accuracy of ±0.5 °C. The Temp1000IS can store up to 32,767 time and date stamped readings, and is able to be programmed to start up to 24 months in advance.
The Temp1000IS is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.
The Temp1000IS has been SGS certified as intrinsically safe for Class I, II, III, Division 1, Groups A-G, -40°C<Tamb< +80°C, T4A and Class I, II, III, Division 2, Groups A-D, F, G, -40°C<Tamb< +80°C, T4A.
| Memory | 65,536 Readings |
|---|---|
| Reading Rate | 1 reading every second up to 1 reading every 24 hours |
| Battery Life | 2 year typical at 15 minute reading rate |
| Material | 316 Stainless Steel/Radel |
| Dimensions | 2.66 in x .97 in dia. (67.6 mm x 24.6 mm dia.) |
| Key Ring End Cap | Key ring replaces standard bottom; Thru-hole: 0.150" |
| IP Rating | IP68 |
| Operating Environment | -40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH (case properly sealed) |
| Required Interface Package | IFC400 |
| Approvals | CE; FM Class 3600, latest revision; FM Class 3610, latest revision; CAN/CSA-C22.2 No. 60079-0:15; CAN/CSA-C22.2 No. 60079-11:14 |
| Technical Rating | Class I, II, III, Division 1, Groups A-G, -40°C<Tamb< +80°C, T4A and; Class I, II, III, Division 2, Groups A-D, F, G, -40°C<Tamb< +80°C, T4A |
| Temperature Sensor | Semiconductor |
|---|---|
| Temperature Range | -40 °C to +80 °C (-40 °F to +176 °F) |
| Temperature Resolution | 0.01 °C (0.018 °F) |
| Temperature Calibrated Accuracy | ±0.5 °C (0 °C to ±55 °C), ±0.9 °F (32 °F to 131 °F) |
Elija modelo, sondas y accesorios. Las combinaciones incompatibles se bloquean solas; el resto de las reglas se verifica al agregar el equipo a la cotización.
Cargando las opciones…
Se puede pegar un código y la configuración se arma sola.
Se piden por separado: no hace falta configurar el equipo.

Magnetic Holder for use with the PR140, PRTemp140, HiTemp140, Temp1000, and RHTemp1000 Series Data Loggers.


Versatile mount or stand for use with the HiTemp140 series of data loggers.


Versatile stand for use with the HiTemp140, PRTemp140, PR140, Temp1000 and RHTemp1000 series of data loggers.

Instrumentos para cotizar
Todavía no hay instrumentos. Se puede escribir igual: un especialista responde.