Medidor de peróxido de hidrogênio vaporizado, umidade e temperatura HPP272
para processos de biodescontaminação
Nota de aplicação
An isolator is a device supplied with air meeting a minimum of ISO 5 classification standards that provides uncompromised, continuous isolation of its interior from the external environment. Isolators are used in a variety of life science applications, such as the manufacture of pharmaceuticals and vaccines, aseptic filling, sterility testing, and pharmacy compounding. Isolators must be treated regularly with disinfection agents in order to ensure they are free of microorganisms.
These agents need to be effective against all possible forms of microbes: viruses, bacteria, yeast, mold, and fungi, including highly resistant forms like mycobacteria and bacterial spores. A minimum 6-log bioburden reduction in microorganisms needs to be achieved when treating an isolator with bio-decontamination agents.
Vaporized hydrogen peroxide (H2O2 ) is a commonly used disinfection agent for this purpose. Vaporized H2O2 is non-toxic, works at low temperatures, and destroys the full spectrum of biological contaminants. It is compatible with a wide variety of materials. Because it decomposes to water vapor and oxygen, it leaves no residue on surfaces, meaning they do not have to be wiped down after bio-decontamination.
It is therefore no surprise that vaporized H2O2 is widely used in place of other agents (chlorine dioxide, formaldehydes, and ethylene oxide) to kill microorganisms.
When performing isolator bio-decontamination cycle development studies, accurate, stable, and repeatable measurement values for H2O2 ppm content, humidity (referring to both relative humidity and relative saturation), temperature and time of exposure are crucial.
Theoretically, reproducible cycle development can be divided into seven separate steps (all steps are not mandatory):
Reliable measurements of environmental conditions and their distribution inside an isolator are needed in order to perform all of these steps and to validate biodecontamination cycles quickly and easily.
Once an isolator’s qualification steps have been performed and the bio-decontamination cycle validated, bio-decontamination with vaporized H2O2 is repeated according to the needs of the manufacturing processes. The bio-decontamination cycle interval depends on the product being manufactured and types of processes being performed in an isolator.
Depending on the process in question, bio-decontamination may be performed as often as several times per day, or as infrequently as once per year. Disinfection is a critical process and often a time consuming one.
Bio-decontamination can be divided into four separate steps, all of which must be carefully controlled and monitored. The four steps (see Figure 1) are described below.
Rigorous control and monitoring of the environmental conditions (H2O2 ppm, RH, RS and temperature) helps guarantee the safety of the products produced in the isolator and decreases the amount of time and money spent during decontamination. The Vaisala PEROXCAP® Hydrogen Peroxide, Humidity, and Temperature Probe HPP270 series fulfills several isolator measurement needs.

Figure 1. An example of the four steps in a typical non-condensing isolator bio-decontamination process.

para processos de biodescontaminação

O mais novo integrante da linha Indigo de transmissores e sondas, o Indigo80 é um coletor de dados e indicador portátil e compacto. Ideal para…

para inspeções rápidas e verificação do local

Transmissores de ±1,5% de U.R. para HVAC exigente, com um certificado

A sonda de umidade HMP113 com gabinete de plástico possui excelente confiabilidade e alta tolerância química, e é adequada para ambientes sem…

para aplicações de umidade e temperatura de finalidade geral com sensor intercambiável

Princípio de funcionamento dos sensores infravermelhos, compensação por temperatura, pressão e umidade, e localização de transmissores.

Por que o CO2 é crítico na cultura celular, o risco de condensação ao medir em alta umidade e temperatura, e por que a esterilização sem remover o…

Por que o O2 é reduzido e o CO2 é elevado em câmaras de atmosfera controlada, e quais instrumentos toleram o frio e a umidade elevada desse ambiente.

O que é a amostragem por equilíbrio, por que o aquecimento do sensor pode alterar a medição em campo, e qual faixa é necessária para perfis profundos…
Um especialista responde com a solução, o instrumento e o serviço adequados.
Instrumentos para cotar
Ainda não há instrumentos. É possível escrever mesmo assim: um especialista responde.