Uma unidade da AKRIBIS One

Gold-Platinum-Thermocouple

Gold / Platinum Thermocouple

Since 1995 Isotech have been producing various designs of special Au/Pt, Pt/Pd, Pd/Au thermocouples for researchers. From our experience we can now offer the most popular of these, the Pt/Au thermocouple in a standard form.

Solicitar cotaçãoBaixar ficha técnica

Gold / Platinum Thermocouple

Features

  • Pure Metal Construction: 99.999% Gold and Platinum
  • Best Homogeneity
  • Economic Alternative to HTSPRTs

Description

All wires are 99.999+% pure and are fully annealed according to the recommendations of McLaren. Assembly also follows his prescriptions which have never been bettered.

After final assembly and annealing the Pt/Au thermocouples will conform to IEC 62460, Edition 1 2008-07.

For the smallest uncertainties we calibrate the thermocouple at the Zinc, Aluminium and Silver Fixed Points.

Specifications

Temperature Range0°C to 1000°C
Sheath Materials Measuring JunctionQuartz
Reference JunctionStainless Steel

Thermo-element purities:

Platinum99.999% pure
Gold99.999% pure
Calibration optionsIsotech traceable calibration at Zinc, Aluminium and silver fixed points. NPL: Fixed Point Calibration: Calibrated at fixed points of Zn, Al and Ag (UKAS)
Uncertainty0-400°C ±0.07°C. 400-1000°C ±0.05°C
DimensionsRefer to diagram
Carrying CaseIncluded as standard
How to orderModel type: Au/Pt Thermocouple. Including emf vs. temperature traceable calibration certificate and carrying case.

Documentos

  • SPRT-108462

    Copper Point SPRT Model 108462

    The ITS-90 specifies the SPRT to the freezing point of silver, 961.78°C. Thermocouples can be used beyond this temperature but it is difficult to see…

  • Model-465

    Three Zone High Temperature Furnace 465

    Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…

  • Low-Temperature-Furnace

    Low Temperature Furnace

    The Isotech Low Temperature Fixed Point Furnace is designed specifically to realize and maintain the freeze plateaux of Isotech Indium, Tin and Zinc…

  • Water-Triple-Point-Cells

    Jarrett-Isotech Water Triple Point Cells

    Internationally proven water triple point standards The Water Triple Point is the most important fixed point in temperature metrology, and the only…

  • Gallium-Fixed-Point-Cell

    Gallium Fixed Point Cell

    Second only to the Water Triple Point and in many ways, because of its ease of use and purity, superior to it, is the Gallium Melt Point. At…

  • SPRT-670

    Isotech Model 670 SPRT

    The 670SQ is our latest thermometer, to be specifically designed to give optimum performance up to the aluminium point. Its construction permits the…