Isotech Model 96178 HTSPRT
Isotech has produced over 200 high temperature thermometers which have been sold world-wide for use up to the silver point. As a consequence of our…
SPRT-108462
The ITS-90 specifies the SPRT to the freezing point of silver, 961.78°C. Thermocouples can be used beyond this temperature but it is difficult to see small changes in temperature. For a Type R thermocouple the voltage sensitivity at the copper point is 14μV/°C; a change of 1 μV is equivalent to a voltage change of 71mK. When using thermocouples electrical noise limits the ability to follow small changes in the copper freezing plateau and so attention was turned to the development of a new resist
Solicitar cotaçãoBaixar ficha técnica

Isotech has a long history of making SPRTs to the Silver point (Model 96178) and this experience was combined with new research to produce the new copper point SPRT (Model 108462).
Ro is nominally 0.25Ω the same as the silver point SPRT but the platinum winding is held in place on a new type of synthetic sapphire mandrel. The platinum ‘loves’ oxidising but ‘hates’ reducing atmospheres. The thermometer sheath is made of alumina. It is air filled and hence surrounded by 20% oxygen. Uniquely the sheath is connected to a small air pump to pressurise the 108462 with air so that any leakage is outwards, whilst maintaining an oxygen rich atmosphere around the winding. This is what gives the thermometer its stability.
The four platinum lead wires are separated with tubes of quartz glass and passed through four bores. In use the winding is biased to +9V DC with the included ioniser.

| Measuring Range | 0 to 1100°C |
|---|---|
| Diameter | 7.5mm |
| Model | 108462 |
| Length | 650mm |
| Ro | 0.25 Ohms |

Isotech has produced over 200 high temperature thermometers which have been sold world-wide for use up to the silver point. As a consequence of our…

Since 1995 Isotech have been producing various designs of special Au/Pt, Pt/Pd, Pd/Au thermocouples for researchers. From our experience we can now…

The Isotech Low Temperature Fixed Point Furnace is designed specifically to realize and maintain the freeze plateaux of Isotech Indium, Tin and Zinc…

Are you fed up with cold wet hands, and hours of frustration when you produce an ice mantle in your Triple Point Cell?

This dedicated apparatus has more than 35 years of proven use and is widely used by the worlds’ leading NMIs. In a recent International study cells…

The 670SQ is our latest thermometer, to be specifically designed to give optimum performance up to the aluminium point. Its construction permits the…

The Isotech Argon Triple Point Apparatus has been upgraded for greater automation, efficiency, and ease of use in realising the argon triple point.

High-Accuracy Reference Thermocouple for Calibration Labs and High-Temperature Applications