QuickCal Blackbody
When the high accuracy of the Gemini R is not necessary this product offers a cost effective solution for the calibration and testing of infrared…
Hyperion-R-982
The Hyperion R Portable Blackbody Calibration Source allows for calibration of non-contact infrared thermometers over the temperature range -20°C to 125°C.
Solicitar cotaçãoBaixar ficha técnica
One application is the calibration of non-contact medical thermometers. With the sudden step increase in the demand to check and calibrate medical thermometers we have supplied many units for both portable and laboratory testing relating to the coronavirus outbreak.
It is suitable for use as a primary radiation source for infrared thermometers from sub zero to 125°C.
Laboratory performance and low uncertainty calibrations are ensured by the combination by high emissivity and excellent temperature uniformity.
The digital temperature controller allows the block temperature to be set to any value from -20°C to 125°C. Traceability of the radiance temperature is established by a separate, built-in temperature indicator and included platinum resistance thermometer.
A three point traceable calibration certificate is included. UKAS calibration of the resistance thermometer is available, as is radiometric calibration.
Uniformity of the block is ensured by using distributed thermoelectric heat pumps with the benefit of solid state, vibration free cooling.
The Hyperion R can be used with the Gallium Fixed Point Cell, Model Number 982-05-01. Melt plateau times of five hours are easily obtainable. The cells are provided with Certificate of Metal Analysis, 6N pure – 99.9999%.

| Model | 982 |
|---|---|
| Temperature Range | -20°C to 125.0°C |
| Emissivity | Greater than 0.995 |
| Stability | ±0.1°C |
| Display Resolution | 0.1°C to 0.01°C |
| Aperture Diameter | 50mm |
| Cavity Depth | 150mm |
| PC Interface | Included |
| Power | 200 Watts typical |
| Voltage | 80-264 Vac 50/60 Hz |
| Dimensions | Height 310mm Width 265mm Depth 200mm |
| Weight | 8.3 kg |
| Options | Gallium Fixed Point Cell 29.7646°C 982-05-01 |
| How to order | Model 982 Hyperion R: State Supply Voltage. Please state any special calibration requirements. |
| Note | Orifice Plates 10, 20, 30, 40mm (Restricts Cavity Aperture) – 812-01-06 |

When the high accuracy of the Gemini R is not necessary this product offers a cost effective solution for the calibration and testing of infrared…

The Gemini R 550 Portable Blackbody Calibration Source allows for calibration of noncontact infrared thermometers over the temperature range 30°C to…

Improves Accuracy of Fever Detection Systems; Can be used either horizontally or vertically.; Check Non-Contact Thermometers; Features a high…

The Oberon R uses a Sodium heat Pipe to ensure an exceptionally low temperature gradient along the furnace core. It is ideal for the realization of…


Primary Standard cells fit into the Medusa and Oberon, covering from Indium to Copper Points. These cylindrical cells are 42mm diameter and have a…
Instrumentos para cotar