ISOTowers-Series

ISOTowers

The most accurately defined temperatures are those defining points (fixed points) of ITS-90.

Solicitar cotaçãoBaixar ficha técnica

ISOTowers

Features

  • High accuracy
  • Fast to temperature
  • Simple to use
ISOTowers

Description

The leading Primary Laboratories use large fixed point cells in deep calibration furnaces that utilise heat pipes to eliminate temperature gradients. This combination of cell and furnace gives the smallest of uncertainties.

In the patented Isothermal Towers the fixed point cell and heat pipe (or heat siphon) have been combined to produce the ideal realisations for calibrating standard thermometers.

Thermometers can only be calibrated accurately if they are immersed sufficiently.

In Isothermal Towers a heated block (Immersion Compensator, patent applied for) sits on top the heat siphon/cell to fully compensate for the immersion characteristics of the unit under test.

The Isothermal Towers performance has been fully evaluated against the most detailed and demanding requirements ever written: CCT/2000-13.

All Isothermal Towers; Indium, Tin, Zinc and Aluminium meet all the requirements of CCT/2000-13 allowing laboratories to realise the smallest uncertainties, at a fraction of the cost of conventional Metrology Furnaces with Primary Standard Cells.

ISOTowers

You can purchase three Isothermal Towers; Tin, Zinc and Aluminium for a similar price as one conventional cell and heat pipe apparatus!

Isothermal Towers are simple to use, and very robust. Operation is risk free, as a combined apparatus there is no need to handle a fragile cell. No need for specialist training courses. Isothermal Towers remove the mystery from fixed point calibration.

Easily set to provide a melt or freeze of 24 hours or more, lending themselves for automatic calibration and providing your lab with an all day long plateau.

Introducing the ISOTowers

ISOTowers

Specifications

Model490491492493
ITS-90 PointIndiumTinZincAluminium
Temperature156.5985°C231.928°C419.527°C660.323°C
Metal Purity6N6N6N6N
Plateau Durationup to 30 hours“““
UKAS Uncertainty: Primary Service±0.6mK±0.8mK±1mK±2mK
UKAS Uncertainty: Standard Service±2mK±2mK±2mK±6mK
Heating Time2 hours“““
Pocket Diameter8mm8mm8mm8mm
Total Immersion Depth290mm290mm290mm290mm
Depth of metal surface to bottom of reentrant tube180mm180mm180mm180mm
Power900W900W900W900W
Voltage110Vac or 230Vac 50/60Hz110Vac or 230Vac 50/60Hz110Vac or 230Vac 50/60Hz110Vac or 230Vac 50/60Hz
DimensionsHeight 430mm, Width 310mm, Depth 300mmHeight 430mm, Width 310mm, Depth 300mmHeight 430mm, Width 310mm, Depth 300mmHeight 430mm, Width 310mm, Depth 300mm
Weight15kg15kg15kg15kg

Configurar o equipamento

Escolha modelo, sondas e acessórios. As combinações incompatíveis são bloqueadas na hora; as demais regras são verificadas ao adicionar o equipamento à cotação.

Documentos

  • Annealing-Furnace

    Annealing Furnace

    The Annealing Furnace, model 414, is designed to heat, anneal and cool Standard Platinum Resistance Thermometers (SPRTs) prior to calibration. The…

  • Medium-Temperature-Furnace

    Medium Temperature Furnace

    Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…

  • Model-471

    Model 471 Simple Argon Apparatus

    The Isotech Argon Triple Point Apparatus has been upgraded for greater automation, efficiency, and ease of use in realising the argon triple point.

  • Model-17402

    Gallium Fixed Point Apparatus

    Model 17402B Gallium Apparatus is not an adaptation from general purpose equipment, but is designed specifically to realize and maintain the Gallium…

  • Gallium-Fixed-Point-Cell

    Gallium Fixed Point Cell

    Second only to the Water Triple Point and in many ways, because of its ease of use and purity, superior to it, is the Gallium Melt Point. At…

  • ITS-90-Optimal-Fixed-Point-Cells

    ITS-90 Optimal Fixed Point Cells

    Isotech Ultra Pure-Metal Freezing Point Cells are specifically designed to realize the liquid-solid equilibrium temperatures of high-purity metal…