Gallium Fixed Point Cell
Second only to the Water Triple Point and in many ways, because of its ease of use and purity, superior to it, is the Gallium Melt Point. At…
Model-1600
High-Accuracy Reference Thermocouple for Calibration Labs and High-Temperature Applications
Solicitar cotaçãoBaixar ficha técnica

The Isotech Model 1600 is a premium Type R or Type S thermocouple standard, designed for calibration laboratories and high-temperature environments up to 1600 °C.
It is ideal for laboratories calibrating with compliance to AMS 2750 and ASTM E2750, including aerospace, automotive, and energy sectors.
Manufactured using ultra-high purity Platinum/Platinum Rhodium wire (Pt/13%Rh for Type R, Pt/10%Rh for Type S), the Model 1600 ensures excellent long-term stability and low drift. A continuous noble metal conductor runs from the hot junction to the cold junction with no intermediate splices, preserving measurement integrity.
The measuring junction is protected inside a gas-tight 7 mm recrystallized alumina ceramic sheath. Inside is a removable 2.5 mm twin-bore insert housing the thermocouple wires. This removable assembly enables independent calibration or replacement without disturbing the outer sheath.
At the cold end, a 250 mm stainless steel sheath provides a stable 0 °C reference junction, connected via 2 m low-EMF copper leads. For systems with internal referencing, a “No Cold Junction” (NCJ) option is available.
| Model | 1600 |
|---|---|
| Hot Sheath Temperature Range | 0°C to 1600°C (R or S) |
| Emf vs Temperature | According to relevant document |
| Response Time | 5 minutes |
| Hot Junction Dimensions | see diagram |
| Connecting Cable | See diagram |
| Cold Junction | 250.0mm long x 4.5 diameter |
| Copper Extension Wires | 2000mm |
| Short Term Stability | ±0.2 deg C or better |
| Immersion | 100mm min. |
| Case Dimensions | Height 65mm, Width 710mm, Depth 165mm, Gross Weight 900g |
| Feature | Removeable inner assembly |
| The standard thermocouple described can be supplied in the following noble metal combinations: | Type R: Platinum v Platinum 13% Rhodium or Type S: Platinum v Platinum 10% Rhodium |
| How to order | Model 1600 type R/300, Model 1600 type R/600, Model 1600 type S/300, Model 1600 type S/600. If cold junction not required, specify NCJ. UKAS Calibration included. |

Second only to the Water Triple Point and in many ways, because of its ease of use and purity, superior to it, is the Gallium Melt Point. At…

The most accurately defined temperatures are those defining points (fixed points) of ITS-90.


The 670SQ is our latest thermometer, to be specifically designed to give optimum performance up to the aluminium point. Its construction permits the…

This dedicated apparatus has more than 35 years of proven use and is widely used by the worlds’ leading NMIs. In a recent International study cells…

The ITS-90 specifies the SPRT to the freezing point of silver, 961.78°C. Thermocouples can be used beyond this temperature but it is difficult to see…

Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…

Whilst heat pipe furnaces offer the ideal environment to melt and freeze ITS-90 Fixed Points the temperature range is limited by fluid that flows…
Instrumentos para cotar