The amount of water in food products, especially in dairy produce, is limited for technological reasons (expected product quality) and legal reasons, for example with regard to butter. Butter is a mixture of milk fats (80-90%), including saturated and unsaturated fatty acids and water. Water is the cheapest ingredient of this mixture so in high-volume production even a small deviation from the water content norm assures conspicuous, yet “sham” savings. It must be noted that the surplus of water in the butter contributes to oxidation of fat (fat goes rancid), which is very unfavorable. It is therefore important to test water content in the butter as part of every engineering process which confirms that legal requirements have been met. The butter water content analysis method must be precise and quick because the test information is often used to control the production. Speed and precision of the analysis are assured by MA R, MA X2, MA X7 and MA 5Y moisture analyzers, manufactured by Radwag, that are also used during final evaluation of butter quality in the KJ Laboratory.

The application note includes basic information for validation of the butter drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.
The method with the use of IR radiation
ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water /dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.
ISO 3727-1. Butter — Determination of moisture, non-fat solids and fat contents — Part 1: Determination of moisture content (Reference method).
Before testing, samples must be stored in sealed packaging at the temperature of 2 o C ÷ 14 o C. Heat the sample dedicated for testing to the temperature between 24 o C and 30 o C.
Dryer, pumice, weighing vessels with a lid, AS 220.X2 balance, laboratory spoon.
Weigh glass vessels with a pre-dried pumice in the amount of ca. 10 g. Place the sample with a mass of ca. 5 g on the pumice and specify the real mass of the sample in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Place weighing vessels with the sample and lids in the temperature-controlled laboratory dryer. Dry samples at the temperature of 102 o C for 2 hours. After this period, remove vessels and place in the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 30 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying. At the same time conduct the so-called “blind” test with the use of the same procedure, yet without the test sample.

Standard deviation (%) 0.15
The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature.
Before testing, samples must be stored in sealed packaging at the temperature of 2 o C ÷ 14 o C. Heat the sample dedicated for testing to the temperature between 24 o C and 30 o C.
MA R, MA X2, MA X7 or MA 5Y moisture analyzer, quartz sand, weighing vessels with a lid, AS 220.X2 balance, laboratory spoon.
Set drying parameters presented below. Place the sample with a mass of ca. 1.5 ÷ 2 g on pre-dried quartz sand in several spots of the weighing pan. Lock the drying chamber manually or automatically.

Analysis time x´ (min) 17

The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.


The Vaisala viewLinc Monitoring System tracks environmental conditions wirelessly using Vaisala’s VaiNet wireless devices based on LoRa®* technology…

One of the most important requirements to reduce cooling costs is to measure conditions properly in the first place. The first things to consider are:

Carbon dioxide and other gases consisting of two or more dissimilar atoms absorb infrared (IR) radiation in a characteristic, unique manner. Such…

Another approach to preventing explosions is to install electrical equipment inside of “explosion proof” housings. These housings can withstand…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.