Application note
The amount of water in food products is one of quality parameters that determine the product flavor and its use-by date. The surplus of water in the product structure is unfavorable because it activates hydrolytic transformations that result in considerable drop of quality and product sensory qualities, such as flavor, aroma, hardness, etc. For this reason the water content check is significant at all stages of the production. On the other hand a precisely measured and limited amount of water is a guarantee of long life of the product, which is one of pro-consumer steps taken by every manufacturer. The method of checking the water content to be used in testing must be accurate and assure highly precise values. This can be achieved in case of using MA R, MA X2, MA X7 and MA 5Y moisture analyzers by Radwag.

The application note includes basic information for validation of the nuts and almonds drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.
The method with the use of IR radiation
ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water / dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.
The reference method parameters are usually specified in standards or other discipline-specific documents as the so-called guides. If such documents are unavailable, the drying temperature that does not cause the sample to change colors is used. Such an approach applies to previously dehydrated products and raw products.
Before testing, samples must be stored in sealed packaging. The samples that take a form of grains (almonds, nuts) must be mechanically fragmented into small pieces. Paste-like samples must be mixed.
Laboratory dryer, glass weighing vessels with a lid, quartz sand, glass rod, AS 220.X2 analytical balance, laboratory spoon.
Place samples of grains with a mass of ca. 5 in pre-dried glass weighing vessels. Mix the semi-fluid (paste) sample with a mass of ca. 3 g and then put into glass weighing vessels on pre-dried quartz sand. Mix the sample with sand by means of a glass rod that must be left in the vessel. The use of sand as a foundation is aimed at eliminating creation of the shell on the surface of the sample in question. Specify the real mass of samples in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Place weighing vessels with the sample and lids into the temperature-controlled laboratory dryer. Dry samples at the temperature of 105 o C for 3 hours. After this time, remove vessels and put into the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 30 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying.

The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature.
Before testing, samples must be stored in sealed packaging. Samples that take a form of grains (almonds, nuts) must be mechanically fragmented into small pieces. Paste-like samples must be mixed.
MA R, MA X2, MA X7 or MA 5Y moisture analyzer, laboratory spoon, disposable aluminum weighing pans.
Set drying parameters presented below. Collect the sample with a mass of ca. 4 ÷ 5 g and distribute a thin layer of the sample throughout the weighing pan. A thin layer of pastes must be distributed throughout the weighing pan. Lock the drying chamber manually or automatically.



The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.

The Vaisala viewLinc Monitoring System tracks environmental conditions wirelessly using Vaisala’s VaiNet wireless devices based on LoRa®* technology…

One of the most important requirements to reduce cooling costs is to measure conditions properly in the first place. The first things to consider are:

Carbon dioxide and other gases consisting of two or more dissimilar atoms absorb infrared (IR) radiation in a characteristic, unique manner. Such…

Another approach to preventing explosions is to install electrical equipment inside of “explosion proof” housings. These housings can withstand…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.