The moisture of harvested rice is more than 20% and it is too much because the surplus of water results in unfavorable changes, e.g. mold development. For this reason rice must be cleaned and dried to reach the moisture of fewer than 14% shortly after the harvest time. For the purposes of rice dehydration, various methods are adopted, for example column dryers, recirculation dryers, fluid dryers, etc. Rice is hygroscopic so the information on water content during storage is important. Keeping a suitable moisture level while storing and transporting rice is required to assure its commercial class and to prevent growth of fermentation. The precise information on rice moisture is therefore required both during processing and storage. Such information can be obtained quickly with the use of MA R, MA X2, MA X7 or MA 5Y moisture analyzers by Radwag. The method of specifying water content of rice has been validated so parameters of the moisture analyzer method guarantee precise results.

The application note includes basic information for validation of the white rice drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.
The method with the use of IR radiation
ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water / dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.
The reference method parameters are usually specified in standards or other discipline-specific documents as the so-called guides. If such documents are unavailable, the drying temperature that does not cause the sample to change colors is used. The water content of rice has been tested as per requirements of the EN ISO 712. Cereals and cereals products. Determination of moisture content. Reference method .
Before testing, the sample must be stored in a tightly sealed container. Before testing, mix and mechanically fragment the sample.
Laboratory dryer, glass weighing vessels with a lid, AS 220.X2 analytical balance, laboratory spoon
Place the sample with a mass of ca. 5 g in pre-dried glass weighing vessels. Specify the real mass of the sample in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Put weighing vessels with the sample and lids in the temperature-controlled laboratory dryer. Dry samples at the temperature of 130 o C for 3 hours. After this period, remove vessels and put into the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 30 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying.

The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature. Too high drying temperature may cause surface burning of the sample, which may be hard to diagnose if the sample color is dark.
Before analyzing, samples must be stored in sealed containers. Mix the sample, grind it down with a grinder, and then collect for testing.
MA R, MA X2, MA X7 or MA 5Y moisture analyzer, laboratory spoon, disposable aluminum weighing pans.
Set drying parameters presented below. Collect the fragmented sample with a mass of ca. 4 g and distribute a thin layer of the sample throughout the weighing pan. Lock the drying chamber manually or automatically.

Analysis time x´ (min) 16

The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.


The Vaisala viewLinc Monitoring System tracks environmental conditions wirelessly using Vaisala’s VaiNet wireless devices based on LoRa®* technology…

One of the most important requirements to reduce cooling costs is to measure conditions properly in the first place. The first things to consider are:

Carbon dioxide and other gases consisting of two or more dissimilar atoms absorb infrared (IR) radiation in a characteristic, unique manner. Such…

Another approach to preventing explosions is to install electrical equipment inside of “explosion proof” housings. These housings can withstand…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.