Application note
While producing cheese spread, a few types of cheese melt together with fluxing agents, water, coloring and flavor additives. In this process, water serves as a plasticizer. Its excessive amount may lead to substantial changes to the product consistency, making it too flexible, or easily deforming, which is unacceptable to consumers. For this reason the amount of water in cheese spread must be inspected during production in a quick and precise way. The dry matter content determination method must allow manipulating the engineering process parameters in order to guarantee top quality of the product. In testing dry matter content in cheese spread, the precise results can be assured by MA R, MA X2, MA X7 and MA 5Y moisture analyzers by Radwag.

The application note includes basic information for validation of the cheese spread drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.
The method with the use of IR radiation
ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water / dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.
The reference method parameters are usually specified in standards or other discipline-specific documents – with regard to the sample in question the guidelines specified in EN ISO 5534:2005. Cheese and processed cheese — Determination of the total solids content (Reference method) have been used.
Before testing, samples must be stored in sealed packaging. Mix the sample if it takes a form of a semi-fluid product. Collect a small amount of the sample for testing. Collect samples from various spots.
Laboratory dryer, weighing vessels, AS 220.X2 balance, laboratory spoon, quartz sand, glass rods.
Weigh glass vessels with a glass rod and pre-dried quartz sand in the amount of ca. 20 g. Place the sample with a mass of ca. 3 g in glass weighing vessels on a pre-dried quartz sand. Stir the sample with sand by means of the glass rod that must be left in the vessel. Use of the sand as a foundation is aimed at eliminating creation of the shell on the surface of the sample in question. Weigh vessels again and specify the real mass of the sample in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Put weighing vessels with the sample and lids in the temperature-controlled laboratory dryer. Dry samples at the temperature of 102 o C for 3 hours. After this period, remove vessels and put into the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 60 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying. Calculate the dry matter content.

The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature.
Before testing, samples must be stored in sealed packaging. Mix the sample if it takes a form of the semi-fluid product. Collect a small amount of samples for testing. Collect samples from various spots.
MA R, MA X2, MA X7 or MA 5Y moisture analyzer, laboratory spoon, disposable aluminum weighing pans.
Set drying parameters presented below. Distribute a thin layer of the sample with a mass of ca. 1.5 ÷ 2 g throughout the weighing pan. Lock the drying chamber manually or automatically to start drying.

Analysis time x´ (min) 14

The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.


The Vaisala viewLinc Monitoring System tracks environmental conditions wirelessly using Vaisala’s VaiNet wireless devices based on LoRa®* technology…

One of the most important requirements to reduce cooling costs is to measure conditions properly in the first place. The first things to consider are:

Carbon dioxide and other gases consisting of two or more dissimilar atoms absorb infrared (IR) radiation in a characteristic, unique manner. Such…