A unit of AKRIBIS One

Application note

Cheese spread – dry matter content determination

Cheese spread – dry matter content determination

Brand
Radwag

While producing cheese spread, a few types of cheese melt together with fluxing agents, water, coloring and flavor additives. In this process, water serves as a plasticizer. Its excessive amount may lead to substantial changes to the product consistency, making it too flexible, or easily deforming, which is unacceptable to consumers. For this reason the amount of water in cheese spread must be inspected during production in a quick and precise way. The dry matter content determination method must allow manipulating the engineering process parameters in order to guarantee top quality of the product. In testing dry matter content in cheese spread, the precise results can be assured by MA R, MA X2, MA X7 and MA 5Y moisture analyzers by Radwag.

The application note includes basic information for validation of the cheese spread drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.

Cheese spread – dry matter content determination

The method with the use of IR radiation

TERMS

ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water / dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.

REFERENCE METHOD

The reference method parameters are usually specified in standards or other discipline-specific documents – with regard to the sample in question the guidelines specified in EN ISO 5534:2005. Cheese and processed cheese — Determination of the total solids content (Reference method) have been used.

SAMPLE PREPARATION

Before testing, samples must be stored in sealed packaging. Mix the sample if it takes a form of a semi-fluid product. Collect a small amount of the sample for testing. Collect samples from various spots.

ACCESSORIES

Laboratory dryer, weighing vessels, AS 220.X2 balance, laboratory spoon, quartz sand, glass rods.

METHOD DESCRIPTION

Weigh glass vessels with a glass rod and pre-dried quartz sand in the amount of ca. 20 g. Place the sample with a mass of ca. 3 g in glass weighing vessels on a pre-dried quartz sand. Stir the sample with sand by means of the glass rod that must be left in the vessel. Use of the sand as a foundation is aimed at eliminating creation of the shell on the surface of the sample in question. Weigh vessels again and specify the real mass of the sample in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Put weighing vessels with the sample and lids in the temperature-controlled laboratory dryer. Dry samples at the temperature of 102 o C for 3 hours. After this period, remove vessels and put into the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 60 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying. Calculate the dry matter content.

RESULTS

DRY MATTER OF THE CHEESE SPREAD DETERMINED WITH THE MOISTURE ANALYZER

The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature.

SAMPLE PREPARATION

Before testing, samples must be stored in sealed packaging. Mix the sample if it takes a form of the semi-fluid product. Collect a small amount of samples for testing. Collect samples from various spots.

ACCESSORIES

MA R, MA X2, MA X7 or MA 5Y moisture analyzer, laboratory spoon, disposable aluminum weighing pans.

METHOD DESCRIPTION

Set drying parameters presented below. Distribute a thin layer of the sample with a mass of ca. 1.5 ÷ 2 g throughout the weighing pan. Lock the drying chamber manually or automatically to start drying.

DRYING PARAMETERS / RESULTS

Analysis time x´ (min)  14

ACCURACY OF THE MA R, MA X2, MA X7, MA 5Y METHOD

RESERVATION

The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.