Food expanding is a physical process during which gas included in the product structure is released as a result of the temperature and pressure differences. The expanded product increases its volume while its structure is “puffed”; it maintains its physicochemical properties by raising its crumbliness, for example wafers. Sensory qualities (flavor) of such products are strictly dependent upon the final water content. The surplus of water in the product structure is unfavorable because it initiates hydrolytical changes that result in substantial drop of the quality and sensory qualities of the product, for example flavor, aroma, hardness, etc. For this reason the water content inspection is essential at all stages of the production. On the other hand a precisely measured and limited amount of water guarantees long lifespan of the product, which is one of every manufacturer’s pro-consumer steps. The method of measuring the water or dry matter content to be used in testing must assure accuracy and highly precise measurements, which can be achieved with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers by Radwag.

The application note includes basic information for validation of the expanded wafers drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.
The method with the use of IR radiation
ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water / dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.
The reference method parameters are usually specified in standards or other discipline-specific documents as the so-called guides. If such documents are unavailable, the drying temperature that does not cause the sample to change colors is used. Such an approach applies to previously dehydrated products and raw products.
Before testing, samples must be stored in sealed packaging. Mechanically fragment wafers into small pieces or powder.
Laboratory dryer, weighing vessels, AS 220.X2 analytical balance, laboratory spoon.
Place the sample with a mass of ca. 3 g in pre-dried glass weighing vessels. Specify the real mass of the sample in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Put weighing vessels with the sample and lids in the temperature-controlled laboratory dryer. Dry samples at the temperature of 105 o C for 3 hours. After this period, remove vessels and put into the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 30 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying.

Standard deviation (%) 0.03 0.02
The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature.
Before testing, samples must be stored in sealed packaging. Mechanically fragment wafers into small pieces or powder.
MA R, MA X2, MA X7 or MA 5Y moisture analyzer, laboratory spoon, disposable aluminum weighing pans.
Set drying parameters presented below. Collect the sample with a mass of ca. 1 ÷ 1.5 g and distribute a thin layer of the sample throughout the weighing pan. Lock the drying chamber manually or automatically.

Analysis time x´ (min) 4 1

The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.


The Vaisala viewLinc Monitoring System tracks environmental conditions wirelessly using Vaisala’s VaiNet wireless devices based on LoRa®* technology…

One of the most important requirements to reduce cooling costs is to measure conditions properly in the first place. The first things to consider are:

Carbon dioxide and other gases consisting of two or more dissimilar atoms absorb infrared (IR) radiation in a characteristic, unique manner. Such…

Another approach to preventing explosions is to install electrical equipment inside of “explosion proof” housings. These housings can withstand…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.