Application note
Wc gel, washing gel, liquid soap – water content determination
The structure of high-quality chemical products contains numerous carefully dosed ingredients thanks to which these products become effective in a relatively diverse environment. What is required is aggressive impact, e.g. WC gel, but also suitable reaction to dirt and water presence, as in a washing gel and liquid soap. The key to designing these phenomena well is surface-active substances, the so-called surfactants. The real amounts of each ingredient added to the mixture must be supervised, which eventually determines the quality and efficiency of the product. One of engineering parameters of numerous chemical product is dry matter, that is a part of the sample that remains after removing all volatile elements. The method of measuring dry matter to be used in testing must guarantee accuracy and highly precise measurements, which can be assured with the use of the MA R, MA X2, MA X7 and MA 5Y moisture analyzers by Radwag.

The application note includes basic information for validation of the WC gel, washing gel, liquid soap drying method with the use of MA R, MA X2, MA X7 and MA 5Y moisture analyzers series by Radwag Wagi Elektroniczne. The application note may be the basis for elaborating own drying method with special regard to distinctive features of the product in question.
The method with the use of IR radiation
ACCURACY of determining water / dry matter content is the difference between the result of the water / dry matter content received in the moisture analyzer method and the result of the water / dry matter content received while drying the same sample through a reference method. PRECISION is a degree of compliance between independent results of the test, received in specific conditions. The measure of precision is a standard deviation from a series of several measurements.
The reference method parameters are usually specified in standards or other discipline-specific documents as the so-called guides. If such documents are unavailable, the drying temperature that does not cause the sample to change colors is used. Such an approach applies to previously dehydrated products and raw products.
Before testing, sample must be stored in sealed containers. Fluid and semi-fluid products must be mixed before testing.
Dryer, quartz sand, weighing vessels with a lid, rods, AS 220.X2 balance, laboratory spoon.
Weigh glass vessels with a glass rod and pre-dried quartz sand in the amount of ca. 15 g. Mix the sample with a mass of ca. 5 g and put into the glass weighing vessels on pre-dried quartz sand. Mix the sample with sand by means of the glass rod that must be left in the vessel. Use of sand as a foundation is aimed at elimination creation of the shell on the surface of the sample in question. Weigh vessels again and specify the real mass of the sample in question with the use of the balance whose weighing accuracy is 0.1 mg (AS 220.X2). Place weighing vessels with the sample and lids in the temperature-controlled laboratory dryer. Dry samples at the temperature of 105 o C for 3 hours. After this period, remove vessels and put into the desiccator until they cool down and weigh afterwards. Place samples in the laboratory dryer again and keep on drying them for 30 minutes. Cool them down and weigh again. Repeat the procedure until you obtain a stable sample mass or record the sample mass growth after drying. Calculate the dry matter content as a quotient of the post-drying sample mass (m 2 ) and pre-drying sample mass (m 1 ).

Standard deviation (%) 0.09 0.04 0.01
The water content testing with the use of the moisture analyzer (IR radiation) entails two phenomena: convection and radiation. The sample temperature rises from outer layers to the bottom of the sample. The temperature gradient in the sample structure minimizes through optimization of the thickness of the dried sample and drying temperature.
Before testing, samples must be stored in sealed packaging. Fluid and semi-fluid products must be mixed before testing.
MA R, MA X2, MA X7 and MA 5Y moisture analyzer, glass weighing vessels with a lid, laboratory spoon, pipette.
Set drying parameters presented below. Distribute a thin layer of the sample with a mass of ca. 1.5 ÷ 2 g throughout the weighing pan. Lock the drying chamber manually or automatically.

Analysis time x´ (min) 9 18 19

The method in question has been verified by the Research Laboratory, yet the results do not include factors arising from diversity of tested samples, operators’ personal skills as well as measuring capability used by moisture analyzer users. For this reason Radwag shall not be held responsible for drying parameters but they can be used to elaborate own drying method.


The Vaisala viewLinc Monitoring System tracks environmental conditions wirelessly using Vaisala’s VaiNet wireless devices based on LoRa®* technology…

One of the most important requirements to reduce cooling costs is to measure conditions properly in the first place. The first things to consider are:

Carbon dioxide and other gases consisting of two or more dissimilar atoms absorb infrared (IR) radiation in a characteristic, unique manner. Such…

Another approach to preventing explosions is to install electrical equipment inside of “explosion proof” housings. These housings can withstand…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.