Autosonde AS41
Upper air soundings provide the crucial data necessary for a complete atmospheric weather profile that leads to accurate forecasting. However…
Technical note


Using viewLinc for non-digital applications
Although digital data is generally preferable in highly regulated environments, viewLinc also has a universal data logger that accepts analog inputs, allowing facilities to use specialized legacy instruments or older equipment that only outputs analog values for parameters that aren’t otherwise offered by Vaisala’s data loggers. This ability to accept almost any measurement is also beneficial where the parameters being measured minimize the error risks associated with loop calibrations or calibrating in the field. A good example of this in viewLinc is differential pressure measurements, which are often used in cleanrooms; high pressure is used in these areas to ensure constant airflow into the dirtier parts of the facility, keeping the area clean by preventing particles from flowing upstream.
Differential pressure measurements are taken in a panel housed near the data logger, allowing for short wires that reduce the impact of wiring on the analog signal and therefore minimize errors. It’s easy to integrate stable, reliable calibration equipment in the field for differential pressure, and this is the exception rather than the rule in this sense. For cleanroom measurements where it’s inconvenient to perform a loop calibration due to longer wires or issues with your calibration reference standard, or where the device signal is too complicated for an analog signal, digital is more helpful. One example of this is particle counters, which are important for determining how clean the room is.
Connecting viewLinc with Modbus
Particle counters provide complicated process data that can’t be easily conveyed with an analog signal and is therefore much better suited to a digital signal. This digital data is ready to travel, so there’s no need to convert it to analog and back again – it just needs a direct digital pathway to the viewLinc system, and the pathway we use is Modbus. To make it easier to convert a Modbus device to viewLinc, we’ve created a library of templates within the system – for example, complicated devices such as particle counters can be added by simply selecting the correct template. It’s also possible to create and save new templates, which you can then use again and again.
Because digital communication remains accurate over long distances, it decreases the number of calibrations needed, increases calibration accuracy, and allows you to simply send the data logger to a calibration lab and replace it with a unit that’s pre-calibrated for your specific process. The viewLinc system has analog sensor capability, but this is only useful in cases where there’s no other option. For this reason, viewLinc uses the Modbus communication protocol wherever possible.
Connecting Modbus Devices to viewLinc 5.1

Upper air soundings provide the crucial data necessary for a complete atmospheric weather profile that leads to accurate forecasting. However…

Vaisala Automatic Weather Station AWS310 is a complete communication and weather observation solution that can be used as a standalone unit, or as…

Our next-generation, cyber secure and modular Maritime Automatic Weather Station AWS830 is engineered for harsh marine environments and compliant…

The Vaisala Automatic Weather Station AWS810 enables modern, high-quality surface weather observation networks anywhere in the world.

Vaisala Maritime Automatic Weather Station AWS830 delivers accurate, always‑on weather, environmental, oceanographic, and motion data for demanding…

Vaisala Beam Weather Station BWS500 is a compact, scalable station for weather, air quality and road condition data, from a single standalone station…

Vaisala DRYCAP® Dew Point and Temperature Probe DMP5 is designed for in-line humidity measurement in industrial drying applications with high…

Dr. Lauri Reuter is an expert in the future of food, and the person to ask if you’ve ever wondered what we are going to eat in the future, and how…

Industrial companies are continuing to take action to minimize their carbon emissions. Many are reducing their use of fossil fuels and switching to…

In the causticizing process, passive chemicals are converted into active cooking chemicals to make white liquor. This is used in the digester as a…
A specialist replies with the solution, the instrument and the service that fit.