Bringing out the beauty of wood: Veneer drying with humidity control
Vaisala DRYCAP® Dew Point and Temperature Probe DMP5 is designed for in-line humidity measurement in industrial drying applications with high…
Technical note
Can I use a Thermocouple Simulator to calibrate my thermocouple based data logger?
Thermocouple calibrators do not work reliably with most MadgeTech thermocouple loggers. The data logger will pull down the voltage of the thermocouple calibrator enough to alter the reading. While the thermocouple calibrator will recover quickly, the data logger will take the reading before the calibrator has corrected itself. This will lead to an inaccurate reading. Due to this, using a thermocouple simulator is not recommended when calibrating MadgeTech loggers.

Vaisala DRYCAP® Dew Point and Temperature Probe DMP5 is designed for in-line humidity measurement in industrial drying applications with high…

Dr. Lauri Reuter is an expert in the future of food, and the person to ask if you’ve ever wondered what we are going to eat in the future, and how…

Industrial companies are continuing to take action to minimize their carbon emissions. Many are reducing their use of fossil fuels and switching to…

In the causticizing process, passive chemicals are converted into active cooking chemicals to make white liquor. This is used in the digester as a…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.