How Often Should a Balance Be Calibrated in a Lab Environment?

Brand
Radwag

As the value of gravitational acceleration varies depending on geographical location, each balance must be adjusted to the location where it will be used to measure mass. Device adjustment is required when it is first installed at its place of use, when its location is changed, and when the ambient temperature changes. To ensure the highest possible weighing accuracy, a correction factor for the balance’s readings relative to a mass standard should be periodically entered into the device’s memory; this is known as balance calibration (adjustment).

Calibration should be carried out:

  • Before starting to weigh
  • If there are long intervals between successive series of measurements
  • When the temperature has changed by more than 3°C.

Types of calibration (adjustment):

  • Automatic internal calibration

Triggered by changes in temperature and the passage of time.

  • Manual internal calibration

Initiated via the balance’s keypad.

  • External calibration

Performed using an external mass standard of a declared mass, which cannot be modified. External adjustment is possible for balances which are not a subject to conformity assessment (verification).

Frequently asked questions

Why does a balance need to be adjusted to its geographical location?

A balance needs to be adjusted to its geographical location because the value of gravitational acceleration varies depending on where the balance is used. This adjustment ensures accurate mass measurements.

When is it necessary to adjust a balance?

Balance adjustment is required when the device is first installed, when its location is changed, or when the ambient temperature changes. Periodic calibration is also recommended for optimal accuracy.

What is balance calibration (adjustment)?

Balance calibration (adjustment) is the process of entering a correction factor into the balance’s memory. This factor relates the balance’s readings to a mass standard, ensuring the highest possible weighing accuracy.

Under what conditions should balance calibration be performed?

Calibration should be performed before starting to weigh, after long intervals between measurements, and when the temperature has changed by more than 3°C.

What are the different types of balance calibration?

The types of calibration are automatic internal calibration, which is triggered by temperature and time changes; manual internal calibration, initiated via the keypad; and external calibration, performed using an external mass standard.

Can external calibration be performed on all balances?

External adjustment is only possible for balances that are not subject to conformity assessment (verification).