Una unidad de AKRIBIS One

RHTemp1000Ex-KR

RHTemp1000Ex (RHTemp1000Ex-KR)

The RHTemp1000Ex humidity and temperature data logger carries hazardous location certification (ATEX, IECEx). Includes ISO 17025 accredited calibration certificate.

Solicitar cotizaciónDescargar ficha técnica

RHTemp1000Ex (RHTemp1000Ex-KR)

Key Features

  • Compact Size
  • ATEX Approved
  • Measures Temperatures from -40 °C to 80 °C

Description

The RHTemp1000Ex is a compact, humidity and temperature data logger that carries hazardous location certification (ATEX, IECEx).

This one of a kind data logger is certified intrinsically safe for the following:

  • Electrical Protection Concepts: ISO/IEC: 60079-11 Ex ia – Ex ic, Intrinsic Safety Zones 0-2
  • Equipment Protection Level: Ga – Gc, Zones 0-2
  • Gas Group: IIC
  • Temperature Class: T4

This device measures and records temperatures from -40 °C to +80 °C (-40 °F to +176 °F) with an accuracy of ±0.5 °C (0 °C to ±50 °C) and ±0.9 °F (32 °F to 122 °F). The RHTemp1000Ex can store up to 32,767 time and date stamped readings, and is able to be programmed to start up to 24 months in advance.

The RHTemp1000Ex is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

General

Memory32,767 readings
Reading Rate1 reading every second up to 1 reading every 24 hours
Battery Life2 years typical at 15 minute reading rate
Material316 Stainless Steel/Radel
Dimensions1.67 in x 0.97 in dia. (42.3 mm x 24.6 mm dia.)
IP Rating(Not Rated) Caution: Do not submerge this product to retain IS rating.
Operating Environment-40 ºC to +80 ºC (-40 °F to +176 °F); 0 %RH to 100 %RH (case properly sealed)
Required Interface PackageIFC400 or IFC406
ApprovalsCE; ATEX Certificate #: 19ATEX0126; IECEx Certificate #: BAS 19.0109

Temperature

Temperature SensorResistance Temperature Detector (RTD)
Temperature Range-40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution0.01 °C (0.018 °F)
Temperature Calibrated Accuracy±0.5 °C (0 °C to ±50 °C), ±0.9 °F (32 °F to 122 °F)

Humidity

Humidity SensorCapacitive Polymer
Humidity Range0 %RH to 100 %RH (non-condensing)
Humidity Resolution0.1 %RH
Humidity Calibrated Accuracy±3 %RH maximum
Humidity Specified Accuracy Range25 %RH to 75 %RH; +20 °C to +40 °C (68 °F to 104 °F); Hysteresis Error 1 % typical, 3 % Maximum

Configurar el equipo

Elija modelo, sondas y accesorios. Las combinaciones incompatibles se bloquean solas; el resto de las reglas se verifica al agregar el equipo a la cotización.

Documentos

  • DLS100

    DLS100

    Versatile stand for use with the HiTemp140, PRTemp140, PR140, Temp1000 and RHTemp1000 series of data loggers.

  • IFC400

    IFC400

    Interface Cable Package for the HiTemp140 and PR140 Data Loggers.

  • IFC406

    IFC406

    IFC406 Multiplexer Data Logger Interface.

  • MagMount-1

    MagMount-1

    Magnetic Holder for use with the PR140, PRTemp140, HiTemp140, Temp1000, and RHTemp1000 Series Data Loggers.

Escribir por WhatsApp