A unit of AKRIBIS One

RHTemp101A

RHTemp101A

The RHTemp101A is a compact, humidity and ambient temperature data logger. Includes ISO17025 accredited calibration certificate.

Request a quoteDownload the datasheet

RHTemp101A

Key Features

  • Compact Size
  • 10-Year Battery Life
  • Programmable Alarm Limits
  • Precision RTD Sensing Element
  • Memory Capacity of 1,000,000+ Readings
RHTemp101A

Description

The RHTemp101A is a compact, humidity and ambient temperature data logger. This device can measure and record temperatures from -40 °C to 80 °C (-40 °F to +176 °F) and humidity from 0 %RH to 95 %RH.

The RHTemp101A features user-specified high/low alarm limits and can even be programmed to start up to 18 months in advance. With a 10-year battery life and the ability to store up to 1,047,552 time and date stamped readings, the RHTemp101A is perfect for long-term studies.

The RHTemp101A is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading simple and easy. Graphical, tabular, and summary data is provided for analysis and can be viewed as customizable engineering units.

General

Memory1,047,552 readings
Reading Rate1 reading every second up to 1 reading every 24 hours
AlarmYes, Humidity
Baud Rate115,200
LED IndicatorsRed & Green
Push-Button StartYes
Battery Life10 years typical at a 15 minute reading rate
MaterialPolycarbonate
Dimensions1.4 in x 2.1 in x 0.6 in (35 mm x 54 mm x 15 mm)
Operating Environment-40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 95 %RH non-condensing
Required Interface PackageIFC200
ApprovalsCE

Temperature

Temperature SensorPrecision RTD Element
Temperature Range-40 °C to +80 °C (-40 °F to +176 °F)
Temperature Resolution0.01 °C (0.018 °F)
Temperature Calibrated Accuracy±0.5 °C (±0.9 °F) from 0 °C to 55 °C (+32 °F to +131 °F); ±1.0 °C (±1.8 °F) from -40 °C to 0 °C (-40 °F to +32 °F) and +55.01 °C to +80 °C (+131.01 °F to +176 °F)

Humidity

Humidity SensorInternal semiconductor
Humidity Range0 %RH to 95 %RH
Humidity Resolution0.1 %RH
Humidity Calibrated Accuracy±3.0 %RH
Humidity Accuracy Range25 % RH to 75 % RH, 20 °C to 40 °C, Hysteresis Error 1 % Typ 3 % Max
RH Units%RH, dew pt., water vapor, concentration (mg/hl)

Configure the instrument

Choose the model, probes and accessories. Incompatible combinations are blocked as you go; the remaining rules are checked when the instrument is added to the quote.

Documents

Write on WhatsApp