Glycol Buffer (Buffer Kit [Bottle, Cap, Seal])
The LogTag® Glycol Buffer is used where the temperature of vaccines needs to be monitored during storage or transport.
UTRED30-16
30 day logger with display, measures and stores up to 30,000 temperature readings
Request a quoteDownload the datasheet

The display is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 30 days. The display also shows the current temperature reading, the current time, recording status and battery status.
The alarm is triggered if temperature readings are outside pre-set alarm limits. The status indicator changes from a ✔ to a X,the red LED is flashing and, if configured, the audible alarm will sound. The day marker will also appear if not already present. The alarm status indicator and audible alarm can be easily cleared, but the day marker remains to ensure a visible record of excursions is shown on the display. A simplified alarm clearance, stop & start procedure is available as an option.
Details of any excursion in the preceding 30 days can be checked directly by inspecting the statistics history on the recorder’s display or in more detail by downloading the logged data via a micro-USB cable to LogTag® Analyzer operating software.

| Product Model | UTRED30-16 |
|---|---|
| External Temperature Sensor Measurement Range | -40 °C to +99 °C. (-40 °F to 210 °F) |
| Recorder Operating Temperature Range | -30 °C to +60 °C 0 – 99 % RH (non-condensing) |
| Storage Temperature Range | -10 °C to +40 °C (32 °F to +104 °F) |
| Rated Temperature Reading Accuracy | Better than ±0.5 °C (±0.9 °F) for -20 °C to +40 °C (-4 °F to +104 °F), typically ±0.3 °C (0.6 °F). Better than ±0.8 °C (±1.4 °F) for -40 °C to -20 °C (-40 °F to -4 °F), typically ±0.5 °C (0.9 °F). Better than ±0.8 °C (±1.4 °F) for +40 °C to +70 °C (+104 °F to +158 °F), typically ±0.7 °C (1.3 °F). Better than ±1.2 °C (±2.2 °F) for +70 °C to +99 °C (+158 °F to +210 °F), typically ±1 °C (1.8 °F) |
| Rated Temperature Reading Resolution | Better than 0.1 °C (0.2 °F) for -30 °C to +60 °C (-22 °F to +140 °F). Better than 0.2 °C (0.4 °F) for -40 °C to -30 °C (-40 °F to -22 °F). Better than 0.5 °C (0.9 °F) for +60 °C to +99 °C (+140 °F to +210 °F). LogTag Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag |
| Sensor Reaction Time | Typically less than 2 minutes(T90) in moving air (1m/s) for ST100S type |
| Recording Capacity | 32,000 total log memory. Provides 16,000 temperature readings per channel. Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics |
| Sampling Interval | Configurable from 30 seconds to 18 hours. (Typically 5 minute logging) |
| Logging Start Options | Push button start or specific date & time. Optional start delay |
| Alarms | Up to 6 individually configurable alarms for each channel. Possible to set alarms separately for Channel 1 and Channel 2 |
| Download Time | Typically with full memory (32,000 readings) in less than 20 seconds |
| Environmental | IP61 (when vertically mounted or hung with USB dust-cap in place) |
| Power source | 5V via micro USB socket or 2 x 1.5V AAA batteries. CR2032 LiMnO2 coin cell (backup logging function only) |
| Battery Life | AAA battery up to 12 months - if unit is operated without power supply |
| Real Time Clock | Built-in real time clock. Typical accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day). Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C) |
| Connection Interface | USB 2.0 / Micro USB Socket |
| Standard Remote Sensor Cable Length | Standard: 1.5m (4’11”). Extended: 3m (9’10”) (recommended maximum) |
| Remote Sensor Cable Type | PTFE (FDA food contact rated) coaxial |
| Size | 141 mm(H) x 72 mm(W) x 18 mm(T) |
| Weight | 156 g with AAA batteries, (132g – excluding probe and AAA batteries) |
| Case Material | Polycarbonate |
| Probe Connectivity | Up to 2 |
| Probe Compatibility | All ST100 precision thermistor external probes can be interchangeably connected to the UTRED30-16. Please refer to the Accessories section for compatible External Probes. Different probe models can be used for each channel. Not compatible with ST10 sensors |
Temperature measurement where access requires a remote probe and data needs to be accessed without a PC. The LogTag UTRED30-16 DUAL provides the facility to create a detailed PDF report without need for extra accessories and proprietary software and features a display, allowing immediate viewing of current and past temperatures. It also has a display so that current and historical temperatures, as well as alarm statistics for the last 30 days, can be viewed instantly without a PC.

Choose the model, probes and accessories. Incompatible combinations are blocked as you go; the remaining rules are checked when the instrument is added to the quote.
Loading the options…
A code can be pasted and the configuration builds itself.
They are quoted separately: the instrument need not be configured.

The LogTag® Glycol Buffer is used where the temperature of vaccines needs to be monitored during storage or transport.


Independently Calibrated Smart Probe. Medium probe, short cable. Compatible with UTRED30-16 Dual Channel, UTRED30-WiFi Dual Channel, TRED30-16U…

Independently Calibrated Smart Probe. Medium probe, long cable. Compatible with UTRED30-16 Dual Channel, UTRED30-WiFi Dual Channel, TRED30-16U. 3.2mm…



Long pointed Tipped Standard Temperature Probe with Handle. Short cable. 5mm Ø × 105mm long. 1.5 m (4'11")

Long pointed Tipped Standard Temperature Probe with Handle. Long cable. 5mm Ø × 105mm long. 3 m (9'10")

Short Tipped Standard Temperature range Probe. Short cable. 3.2mm Ø × 40mm long. 1.5 m (4'11")

Short Tipped Standard Temperature range Probe. Long cable. 3.2mm Ø × 40mm long. 3 m (9'10")

Medium round Tipped Standard Temperature range Probe. Short cable. 3.2mm Ø × 65mm long. 1.5 m (4'11")

Medium round Tipped Standard Temperature range Probe. Long cable. 3.2mm Ø × 65mm long. 3 m (9'10")

Long, round Tipped Standard Temperature range Probe. Short cable. 3.2mm Ø × 90mm long. 1.5 m (4'11")

Long, round Tipped Standard Temperature range Probe. Long cable. 3.2mm Ø × 90mm long. 3 m (9'10")

Short Round Tipped Standard Temperature range Probe. Short cable. 5mm Ø × 22.5mm long. 1.5 m (4'11")

Short Round Tipped Standard Temperature range Probe. Long cable. 5mm Ø × 22.5mm long. 3 m (9'10")

Long, pointed Tipped Standard Temperature range Probe. Short cable. 5mm Ø × 137.5mm long. 1.5 m (4'11")

Long, pointed Tipped Standard Temperature range Probe. Long cable. 5mm Ø × 137.5mm long. 3 m (9'10")

Extension Cable for All External Sensors (excluding Smart Probes). Temperature range -46 °C to +74 °C (-50.8 °F to 165.2 °F). 5 m (16'5")
Instruments to quote