Multi Junction Laboratory Probes
Probes are suitable for use with all Isotech thermocouple reference units or may be used with any other equipment including ice flasks and ice point…
Isorac-844
The function of this reference unit differs from other rack mounted systems in that the temperature of the metal block in which the thermocouple and copper leads are inserted actually varies with ambient temperature. A separate output signal is produced which is proportional to the difference between the environmental temperature and the actual reference temperature. The output signal can be fed directly into a computer/data logger to give accurate compensation over a large ambient range. A ther
Request a quoteDownload the datasheet

The reservoir temperature slowly follows the ambient temperature; an electrical compensation device is thermally integrated with the reservoir and thus senses the reservoir temperature.
The device produces an output proportional to the difference between the reservoir temperature and the reference temperature (usually 0°C). This is the signal the computer/data logger uses to compensate for the temperature of the reference probes junctions.
The output signal can be in the form of a DC mV output, 4-20 mA or from a platinum resistance thermometer. Please discuss prior to order.

| Reference Temperature | Effectively 0°C |
|---|---|
| Temperature Gradient | ±0.1°C between Junctions |
| Stability | ±0.05°C per 1000 Hours |
| Ambient Range | -20°C to +70°C |
| Accuracy | ±0.1°C per 15°C ambient span |
| Stabilisation Time | 10 mins |
| Thermocouple Capacity | 0 to 100 channels double junction referencing |
| Power | 10 Watts typical 100-130 or 208-240 VAC 50/60Hz Low level D.C. Consumption 6VA typical Passive Option Available. |
| Dimensions – Up to 50 Channels | Height 255mm Width 483mm Depth 312mm |
| Weight | 17.2kg |
| 50 to 100 channels | Height 309mm Width 483mm Depth 312mm |
| Weight | 20.4kg |
| How to order | Model 844 – Isoracs are normally uniquely specified for each order. Please discuss your exact requirements with us before ordering. |
| Note | Numerous special versions are available and can be supplied either in their existing forms or modified to customers’ requirements. |

Probes are suitable for use with all Isotech thermocouple reference units or may be used with any other equipment including ice flasks and ice point…

The Hotbox is housed in a robust waterproof enclosure to IP66 incorporating bottom gland plate. Fixing lugs for wall mounting are provided. Easy…

The TRUrac is a 0°C or elevated temperature thermocouple reference system mounted in a 19” chassis. It has been developed for situations where…

The TRU (Thermocouple Reference Unit) supplies a stable and accurate 0°C, or elevated reference temperature. It is a self-contained all solid-state…

The function of this reference unit differs from other cased systems in that the temperature of the metal block in which the thermocouple and copper…
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.