TRU Model 938
The TRU (Thermocouple Reference Unit) supplies a stable and accurate 0°C, or elevated reference temperature. It is a self-contained all solid-state…
TRU2-922
Why use an external cold junction reference?
Request a quoteDownload the datasheet

Internal cold junction compensation in thermometers, scanners and data loggers is often the dominant source of uncertainty, commonly in the range of 0.5°C to 1.0°C and sometimes higher.
TRU² provides a stable, uniform and traceable reference temperature, typically enabling cold junction uncertainties better than ±0.1°C. This simplifies uncertainty budgets, improves confidence, and allows the full performance of precision EMF measurement instruments to be realised.
TRU² is available in three configurations to suit different channel counts and laboratory layouts.
TRU² (base unit)
Direct access to the reference well without a connector panel.
TRU² 50
Front panel fitted, supporting up to 50 double junction thermocouples.
TRU² 100
Front and rear panels fitted, supporting up to 100 double junction thermocouples

TRU² (base unit) provides the 0°C reference block only, with no connector panels fitted. TRU² 50 can be supplied with customer-specified connections on one side of the unit, supporting up to 50 double thermocouple junctions. TRU² 100 provides connections on both sides, supporting up to 100 junctions. Connector type and layout can be customised to suit individual laboratory or bench requirements.
• Screw terminals, suitable for 0.2 mm² to 1.5 mm² thermocouple wire (approximately 24 to 16 AWG)
• Optional miniature or full-size thermocouple connectors
• Connector type specified to suit thermocouple type (K, N, T, etc.)
These options allow direct connection without adapters, reducing unwanted junctions and simplifying laboratory wiring.

The flexible termination and configuration options of TRU² make it suitable for a wide range of calibration and measurement environments, including ISO 17025 accredited laboratories. From high-channel laboratory systems to compact bench installations, TRU² can be configured to match existing wiring practices, reduce unnecessary junctions, and integrate cleanly into both manual and automated setups.
• Accredited calibration laboratories
• National metrology institutes
• Instrument and sensor manufacturers
• Industrial calibration and QA facilities
• Research and development laboratories

| Operating Temperature | 0°C (or 45° to 70°C) |
|---|---|
| Stability | ±0.03°C |
| Capacity | 6 x 8 mm Pockets 195 mm deep alternative / custom inserts available |
| Alarm Facilities | Non-latching relay rated 5 Amps 240V |
| PC Interface | Serial |
| Power | 100 Watts typical 85-264 VAC 50/60Hz |
| Dimensions | Height 338mm Width 325mm Depth 150mm TRU 2 50 150mm TRU 2 100 280mm |
| Weight | TBC |
| Accessories | 935-14-61 – Platinum Resistance Thermometer UKAS Calibration Available |

The TRU (Thermocouple Reference Unit) supplies a stable and accurate 0°C, or elevated reference temperature. It is a self-contained all solid-state…

The function of this reference unit differs from other cased systems in that the temperature of the metal block in which the thermocouple and copper…

The TRU 937 (Temperature Reference Unit) supplies a stable and accurate 0°C or elevated Reference Temperatures between 45°C and 70°C. It is a…

The TRUrac is a 0°C or elevated temperature thermocouple reference system mounted in a 19” chassis. It has been developed for situations where…

Probes are suitable for use with all Isotech thermocouple reference units or may be used with any other equipment including ice flasks and ice point…

The Hotbox is housed in a robust waterproof enclosure to IP66 incorporating bottom gland plate. Fixing lugs for wall mounting are provided. Easy…
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.