IFC200
USB Interface Cable Package for Standard Series Data Loggers.
Technical note
Stereo cables have black or colored plastic insulating rings that separate a stereo cable jack into three separate segments, called the tip, ring, and sleeve.
These three segments are then attached to three wires that are electrically separated with insulation, allowing each channel to send its own signal. Through flexing of the cable over time, the insulating material can split or separate and expose the bare wire of the three channels. This exposure can result in crosstalk, which will interrupt the communication between the IFC200 and data loggers.

It is also possible that the metal of the wire has separated from the jack itself, or separated somewhere inside of the length of cable. This would cause complete communication loss along a single segment.
A multimeter set to measure resistance in Ohms (Ω) can be used to test if either of these issues have occurred. To test for continuity and crosstalk, connect the test leads of the multimeter to each end of the stereo cable on the identical opposing segment.
When testing each segment, there should be a trivial amount of resistance measured from opposing tips, rings, and sleeves respectively. However, there should be no measured resistance between tip-to-ring, tip-to-sleeve, or ring-to-sleeve. If there is any measured resistance between these segments, or if there is no measured resistance between one of the three opposing segments of the cable, then it would be advisable to replace the stereo cable entirely.
When replacing the stereo cable, it’s important to make sure the replacement jack is identical to the original, in that it has three metal segments separated by two insulating rings.

USB Interface Cable Package for Standard Series Data Loggers.

2 position, pluggable terminal block for use with MadgeTech data loggers.

3 position pluggable terminal block for use with MadgeTech data loggers.

4 position pluggable terminal block for use with MadgeTech data loggers.

The EggTemp-RH is an egg-shaped temperature and humidity data logger. Includes ISO17025 accredited calibration certificate.

The EggTemp is an egg-shaped temperature data logger. Includes ISO17025 accredited calibration certificate.

Vaisala DRYCAP® Dew Point and Temperature Probe DMP5 is designed for in-line humidity measurement in industrial drying applications with high…

Dr. Lauri Reuter is an expert in the future of food, and the person to ask if you’ve ever wondered what we are going to eat in the future, and how…

Industrial companies are continuing to take action to minimize their carbon emissions. Many are reducing their use of fossil fuels and switching to…

In the causticizing process, passive chemicals are converted into active cooking chemicals to make white liquor. This is used in the digester as a…
A specialist replies with the solution, the instrument and the service that fit.
Products to quote
No products yet. An enquiry can be sent anyway: a specialist answers.