DMP80 DRYCAP® handheld probe for dew point and temperature
For spot-checking and calibrating dew point measurements
Application note
Compressed air is everywhere in an industrial plant: it drives valves and actuators, conveys and blows product, and often comes into direct contact with the process. The moisture it carries is its silent problem — it condenses in the lines, corrodes piping and equipment, freezes valves in outdoor installations and promotes bacterial growth. For the general context of the magnitude, see the dew point instrumentation landing.
Since compressed air runs dry by design, relative humidity stops being a useful variable: the correct control variable is pressure dew point. The full reasoning, with the compressed air scenario worked through step by step, is in the technical note on practical humidity measurement.
Dew point indicates the temperature at which the water vapor in compressed air condenses. As long as that value stays below the lowest temperature the installation encounters, condensation is impossible anywhere in the system. That is why ISO 8573-1, the standard that classifies compressed air purity, expresses water content as pressure dew point classes — not as relative humidity.
Dryers determine that value. A refrigerant dryer brings the air to dew points around +3 °C, enough for the standard's intermediate classes; the driest classes require a desiccant dryer, which reaches −40 °C or lower. Class 1 — the most demanding — specifies a pressure dew point of −70 °C or below.
Dew point depends on pressure: the same air, expanded to atmospheric pressure, shows a dew point tens of degrees lower than at line pressure. The first installation criterion is therefore to define at which pressure the reading is needed — the standard classifies by pressure dew point — and measure at that condition, with no uncontrolled expansions along the way.
Direct in-line measurement is the simplest option when conditions allow it: DRYCAP® dew point transmitters install directly into pressurized systems. When the gas temperature is high, or the sensor needs protection from water spikes — typical during system failures or start-up —, a sampling cell is inserted to condition the gas before the sensor. In both cases one basic rule applies: the process dew point must stay below the room's ambient temperature, or vapor condenses in the measurement line itself and the reading becomes meaningless.
Continuous dew point measurement turns the dryer into a verifiable piece of equipment: a sustained drift toward wetter values reveals exhausted desiccant, an undersized dryer or an incipient failure — before water shows up in the lines or product gets rejected. It also prevents the opposite mistake: over-drying, which consumes energy for no benefit.
DRYCAP® technology is built for this duty: it withstands condensation and recovers undamaged after getting wet, resists oil vapor and particulates, and its auto-calibration function corrects dry-end drift, with typical calibration intervals of two years.
This technical content is based on Vaisala's documentation on dew point measurement in compressed air. AKRIBIS is an authorized Vaisala partner for distribution and technical support in the region — explore the full Vaisala instrumentation line. For your system's specific configuration — direct measurement or a sampling cell —, consult a specialist.
For spot-checking and calibrating dew point measurements

Please note that the DMT152 is being discontinued and should not be specified for new projects. Get to know the new DMT153 here.

How infrared sensors work, temperature/pressure/humidity compensation, and transmitter placement.

Why CO2 is critical in cell culture, the risk of condensation when measuring in high humidity and temperature, and why sterilization without removing…

Why O2 is lowered and CO2 raised in controlled-atmosphere rooms, and which instruments withstand the cold and high humidity of this environment.

What equilibrium sampling is, why sensor heating can skew field measurements, and what range is needed for deep soil profiles.
A specialist replies with the solution, the instrument and the service that fit.
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.