Vaporized Hydrogen Peroxide, Humidity and Temperature Measurement HPP270 Series
for bio-decontamination processes
Technical note
The PEROXCAP® sensor for vaporized hydrogen peroxide bases its measurement on two HUMICAP® sensors. To understand how HUMICAP sensors work, it is useful to know a little about thin-film polymer sensors. In this type of sensor, there is a thin layer of polymer between two electrodes. The film absorbs or releases vapor according to humidity changes in the environment.
As the humidity changes, the dielectric properties and capacitance of the sensor changes. A dielectric is an insulator that impedes an electric charge; capacitance is the ability of conductive materials to respond to voltage changes. Essentially a thin film polymer sensor is measuring voltage changes that occur with the amount of water vapor in the environment. Electronics within the instrument use the capacitance of the sensor to give a humidity measurement.
The PEROXCAP sensor uses two HUMICAP sensors: one with and one without a catalytic layer. The catalytic layer breaks down the hydrogen peroxide so that the HUMICAP sensor with the layer senses only the humidity, while the sensor without the catalytic layer senses both hydrogen peroxide vapor and water vapor in the air. The instrument calculates the difference between the two sensors to give a measurement of H2O2 concentration.
However, humidity sensors are different from other sensors, such as temperature sensors, because they are in direct contact with the environment they measure. Although the HPP270 series probes provide the highest accuracy and precision, even the best sensors will drift over time. Dirt, chemicals and temperature changes can also cause a drift in accuracy. Regular calibration mitigates incremental sensor drift and ensures instruments are operating within their specifications.
Factory Calibration:
We recommend H2O2 calibration at a Vaisala laboratory for traceable calibration service. Calibration for the hydrogen peroxide measurement uses two different H2O2 vapor concentrations. Calibrations on the HPP272 probes are performed for H2O2, Relative Saturation, Relative Humidity, Temperature and analog output. Note that the HPP271 probes measure H2O2 only (no temperature or relative humidity calibration needed.) Calibration service comes with a certificate and full instrument maintenance is optional. For factory service, find your nearest calibration center.
On-site Calibration:
For on-site calibrations, we offer a do-it-yourself option with the HMK15 RH calibrator as a reference and our Insight software. The software shows you the drift in both Relative Saturation (RS) and Relative Humidity (RH) and gives you the possibility to make two-point adjustment to these parameters in order to show the same value as a reference value. Based on the humidity calibration, the software also calculates H2O2 ppm error at a certain ppm level. High quality H2O2 measurement performance is also ensured by calibration and adjustment. Learn more about Vaisala calibration service.

for bio-decontamination processes

The latest addition to the Indigo family of transmitters and probes, the Indigo80 is a compact, handheld indicator and data logger. Ideal for field…

for quick inspections and spot-checking

For high precision measurements in demanding HVAC applications

The Vaisala HUMICAP® Humidity and Temperature Probe HMP113 is a highly accurate and cost-effective humidity probe with a plastic enclosure. It has…

for general purpose humidity and temperature applications with exchangeable sensor

What relative humidity, dew point, absolute humidity and mixing ratio are, how temperature and pressure distort the reading, and why dew point is…

How humidity behaves in real conditions, the three most common measurement problems, and two practical scenarios —heated air and compressed air—…
When to measure relative humidity vs. dew point, what high and low humidity demand, how temperature and pressure factor in, and when a warmed probe…

Cooling accounts for up to 40% of a data center's energy demand. This Vaisala guide explains how measurement accuracy cuts costs and risks, with a…
A specialist replies with the solution, the instrument and the service that fit.
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.