Application note
Application note
Vaporized hydrogen peroxide (VHP) is a common decontamination agent in life science incubators, isolators, cleanrooms and process lines: it operates at low temperature, is compatible with most materials, and decomposes into water and oxygen with no toxic residue. For the general context of why humidity is controlled in cleanrooms, see the note on humidity instrumentation.
Precise humidity control is critical throughout the VHP cycle: not just for effective decontamination, but to prevent unwanted condensation on surfaces and equipment.
A typical VHP cycle has four phases. In dehumidification, the environment is dried before vapor injection to prevent unwanted condensation. In conditioning, the correct environmental conditions are established before introducing the peroxide. In the bio-decontamination phase itself, target exposure runs between 300 and 1200 ppm of H2O2 with humidity between 50% and 100%. Finally, in aeration, a catalytic conversion breaks down the remaining H2O2 into water vapor and oxygen.
Every phase requires a reliable humidity reading — a measurement error in the wrong phase can mean incomplete decontamination or condensation that damages sensitive equipment.
Relative humidity, by definition, only indicates the level of water vapor in the air relative to temperature. But in a hydrogen peroxide-rich environment, condensation occurs before reaching 100% RH, because both water and peroxide molecules affect the combined saturation point. The variable that does describe that behavior is relative saturation (RS), which depends on the combined concentration of water and peroxide vapor plus air temperature.
A standard humidity sensor, without a protective catalytic layer, can't distinguish between the two vapors: hydrogen peroxide produces a stronger sensor response than water vapor alone, creating measurement errors — on the order of +1% RH at 300 ppm VHP and up to +3% RH at 900 ppm (at 23°C). Sustained exposure to harmful concentrations also causes that type of sensor to drift over time.
A first solution is catalytic HUMICAP® sensors, which incorporate a protective layer that breaks down vH2O2 before it reaches the sensing element — allowing accurate relative humidity measurement between cycles, though they can't determine actual relative saturation during decontamination itself.
Vaisala's PEROXCAP® technology combines a catalytic and a non-catalytic sensor in a single probe, enabling simultaneous calculation of relative humidity, relative saturation and hydrogen peroxide concentration through a proprietary algorithm — with heating and chemical purge included as standard, rather than optional.
This technical content is based on Vaisala's application note on humidity measurement in vaporized hydrogen peroxide bio-decontamination processes. AKRIBIS is an authorized Vaisala partner for distribution and technical support in the region — explore the full Vaisala instrumentation line. For the specific probe for your VHP cycle, consult a specialist.


for monitoring temperature, humidity, with optional analog input

How infrared sensors work, temperature/pressure/humidity compensation, and transmitter placement.

Why CO2 is critical in cell culture, the risk of condensation when measuring in high humidity and temperature, and why sterilization without removing…

Why O2 is lowered and CO2 raised in controlled-atmosphere rooms, and which instruments withstand the cold and high humidity of this environment.

What equilibrium sampling is, why sensor heating can skew field measurements, and what range is needed for deep soil profiles.
A specialist replies with the solution, the instrument and the service that fit.
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.