During one of our webinars on Environmental Mapping Qualification, we received a question from a participant on the difference between mapping temperature only, and temperature with RH. Senior Regulatory expert Paul Daniel answers below:
Dear Paul,
I recently attended your webinar on Mapping Protocols and I wonder – can I use the same methodology you discuss for temperature and humidity mapping studies as with temperature-only mapping studies? Thanks!
J
Dear J,
Short answer: Absolutely, the same methodology would be used for humidity mapping. Almost everything would be the same when considering the protocol document, and the attachments you would use for mapping humidity. Long answer: there are two critical differences in humidity mapping; those are sensor calibration and sensor placement.
Calibration: It is rare that a pre-study or post-study calibration can be done for a humidity sensor unless you have a well-equipped calibration lab. In these cases, there would be a single check before the study to verify that calibration certificates were available for your humidity sensors, and that each sensor was within its calibration interval at the time of use.
Sensor placement: Generally, the easiest practice is to use the same number of humidity and temperature sensors. However sometimes it can make sense to utilize fewer humidity sensors. If you do use fewer humidity sensors, it is important that you understand humidity well enough to explain to an auditor what the expected humidity values would be (based on the temperature) in the spots that had only temperature sensors.
Our Humidity Calculator is a tool that can help with that. The free Vaisala Humidity Calculator allows you to calculate several humidity parameters from one known value. Make unit conversions on the fly, and see the effects of changing ambient conditions, like temperature and pressure. No more installing software for offline use - just bookmark the calculator page or add it to the home screen on your mobile device.
Learn more about Indigo80, Vaisala's handheld device for spot checking.

for quick inspections and spot-checking

For high precision measurements in demanding HVAC applications

The Vaisala HUMICAP® Humidity and Temperature Probe HMP113 is a highly accurate and cost-effective humidity probe with a plastic enclosure. It has…

for general purpose humidity and temperature applications with exchangeable sensor

for high humidities

For use up to zone 0 and 20

What relative humidity, dew point, absolute humidity and mixing ratio are, how temperature and pressure distort the reading, and why dew point is…

How humidity behaves in real conditions, the three most common measurement problems, and two practical scenarios —heated air and compressed air—…
When to measure relative humidity vs. dew point, what high and low humidity demand, how temperature and pressure factor in, and when a warmed probe…

Cooling accounts for up to 40% of a data center's energy demand. This Vaisala guide explains how measurement accuracy cuts costs and risks, with a…
A specialist replies with the solution, the instrument and the service that fit.
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.