Recently, I received a question from an attendee of our webinar "Mastering Temperature Mapping."
Dear Paul,
For small chambers (upright freezers or incubators), we qualify the unit with empty and max load studies. To meet the minimum requirement, we place 10 temperature sensors within the chamber—one in each corner, one in the center, and one near the control probe.
But, do we need to redo the max load study if the shelf configuration inside the chamber changes, for example, changing from five shelves to four?”
My short answer is... maybe.
The original sensor placement, as described, doesn’t account for the shelves themselves. If the shelves are wire racks, airflow may not be significantly impacted. But it’s not just about the shelves—it’s about the product sitting on those shelves.
Here are a few key questions I would answer in deciding whether to requalify:
Let’s break it down by application.
Freezers generally have wider temperature tolerances and often rely on conduction more than airflow. Most have wire shelves and tend to perform better when more thermal mass is present (i.e., more product). A slight change in shelf configuration is less likely to impact performance significantly.
Incubators, on the other hand, usually have much narrower temperature ranges and rely heavily on airflow for thermal uniformity. They often have solid shelves, which can restrict that airflow, making shelf placement and loading patterns more critical. Any change in configuration or loading could affect performance and may warrant requalification.
This question is a good one—and it merits investigation, particularly for incubators. Before launching into re-mapping, I would evaluate whether the changes affect load distribution, airflow, or thermal uniformity. Start by reviewing your risk assessment and qualification rationale, and look closely at how the equipment operates under the new configuration.

for cleanrooms and industrial settings

for refrigerant dryers

for OEM applications

for quick inspections and spot-checking

For high precision measurements in demanding HVAC applications

Vaisala Humidity Calibrator HMK15 makes calibration and spot checking of humidity probes and transmitters easy and reliable, although for critical…

What relative humidity, dew point, absolute humidity and mixing ratio are, how temperature and pressure distort the reading, and why dew point is…

How humidity behaves in real conditions, the three most common measurement problems, and two practical scenarios —heated air and compressed air—…
When to measure relative humidity vs. dew point, what high and low humidity demand, how temperature and pressure factor in, and when a warmed probe…

Cooling accounts for up to 40% of a data center's energy demand. This Vaisala guide explains how measurement accuracy cuts costs and risks, with a…
A specialist replies with the solution, the instrument and the service that fit.
Instruments to quote
No instruments yet. An enquiry can be sent anyway: a specialist answers.